Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The design dispersion at the screen is 111.7mm which means the calibration should be 10.0 GeV / 111.7 mm = 89.5 keV/um. The difference between the measured and expected could be due to the screen calibration. No calibration images ever appear in the elog. The calibration starts at 17.94 um/px when the camera first appears on one of the FACET-II camera server. The calibration then changes to 9.91 um/px starting with this elog entry, but no calibration images appear in the e-log, this value as has been used for all of FACET-II. The value matches the value used at FACET, most likely the old FACET value was used without every taking a new calibration.

...