Identification
Firmware | xxxx | updated date | |
---|---|---|---|
Module identifiers | |||
Nominal Operating Voltages and Currents
Performance Measures
Bad Switch Status
Jungfrau bad gain mode switch status
exp:run (date) |
---|
Medium Mode
cxilv1019 : 142
(11/7/2021)
4193737 pixels in Normal
17 pixels in Med
550 pixels in Low
28 pixels in Normal
4194274 pixels in Med
2 pixels in Low
22 pixels in Normal
3 pixels in Med
4194273 pixels in Low
cxilu7619 : 168
(10/20/2021)
4194254 pixels in Normal
20 pixels in Med
30 pixels in Low
29 pixels in Normal
4194273 pixels in Med
2 pixels in Low
23 pixels in Normal
3 pixels in Med
4194272 pixels in Low
Number of bad pixels (total) | Number of bad switch pixels | |
---|---|---|
cxix53120 : 35 (03/20/2022) | 4213, map | 2066, map |
cxilx5920 : 159 (03/07/2022) | 4564, map | 2067, map |
cxilw6120 : 207 (02/20/2022) | 4499, map | 1955, map |
cxic00120 : 42 (02/11/2022) | 3782, map | 1814, map |
cxic00120 : 2 (01/11/2022) | 857, map | 664, map |
Pedestal
Method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxxx.
Summary Values
Date | Mean | StdDev | #Bad |
---|---|---|---|
Pedestal Map
Noise map
Gain and uniformity (photon source)L
Linearity (signal injection)
Thermal measurements (all thermometers)
Experiment History
Experiment | Hutch | Shifts | Date Deployed |
---|---|---|---|
Commissioning | MFX | 1 | 15 June 2020 |
Calibration Records
Date | Hutch | Date | Record link | Comment |
---|---|---|---|---|
Commissioning | MFX | 7/29/20 | 15 June 2020 |