<calibrated-data> = (<raw-data> - pedestalspedestal) * <gain-factor>

where pedestals and <gain-factor> are per-pixel map combined from 7 gain mode maps.

Calibration

...

procedure

Accumulate dark data and evaluate pedestals for pedestal for five gain modes:
  • 5 calib-cycles: FH, FM, FL, AHL-H, AML-M (gain is not switching in dark)
     and 2x49 charge injection samples (see plots) where gain is switching:
  • 7X7 calib-cycles: AHL
  • 7X7 calib-cycles: AML

...

Then, in current calibration procedure we evaluate pedestals pedestal for two leftover gain modes AHL-L, AML-L as

...

Panel 2(counting from 0) /reg/g/psdm/detector/gains/epix10k/panels/0000000001-0175233793-0553648150-0743498629-0017156647-0000000000-0000000000

...

Pedestal for fixed gain modes FH, FM, FL

...

Pedestal for switching gain modes AHL-H and derived AHL-L               AML-M and derived AML-L

 

...

Offset (B) for switching gain modes AHL-H, AHL-L, AML-M, AML-L

Offsets difference  H(or M) - L for  AHL, AML

...

Gains (G) for switching gain modes AHL-H, AHL-L, AML-M, AML-L

 

...

Mean values

Fixed gain modeFHFMFL
pedestal339232933277
Switching gain modeAHL-HAHL-LAML-MAML-L
pedestal H(M)-from dark, L-evaluated3332182532952601
offset (B) from fit4584307736792985
offset (B) difference H(M)-L1507694
gain (G) from fit470.5215.40.49

Denote

  • p - pedestal
  • B - fit base-level offset parameter, <intensityADU> = B + G * <event-index-from-0-pulser>
  • G - gain

Current (plots above and results in table) naive version of L-pedestals evaluation

  • pL = pH - (BH-BL)

Gabriel's version

  • pL = BL - (BH-pH)*GL/GH

Image Added

L-pedestals using Gabriel's formulae for AHL-L and AML-L

Image AddedImage Added

Questions

  • do we need to account for the difference between offset (B) and pedestal (Gabriel's version)
  • the same difference may be between fit B(L) and pedestal(L) OR it is smaller due to the gain? 

  • is current procedure correct?
  • if not,
    • what is correct procedure?
    • or what could we calibrate with charge injection?

2020-06-22 ASIC issue

detdaq18 run 52

Panel 14 (counting from 0) /reg/g/psdm/detector/gains/epix10k/panels/0000000001-0175696641-1056964630-1010951301-0019435010-0000000000-0000000000

Data for a few charge injection points

Image AddedImage AddedImage Added

Injected charge or gain is too small to switch to L-gain mode...

 

Gains AHL-H, AHL-L, AML-M, AML-L

Image AddedImage AddedImage AddedImage Added

Panel 14(next-to-last) charge injection in ASIC1 does not work

2020-06-24 calibration constants for ALL gain ranges and panels 

detdaq18 run 52

pL were evaluated using Gabriel's formulae.

plots for pedestals, pixel_rms, pixel_status, pixel_gain (default), gain from charge injection

Image AddedImage AddedImage Added

Image AddedImage Added

deployed as

/reg/d/psdm/det/detdaq18/calib/Epix10ka2M::CalibV1/DetLab.0:Epix10ka2M.0/*/52-end.data

SHAPE  (7,16,352,384) : 7-gains, 16-panels of shape (352,384)

References

...