Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Use the ePixHrM designated power BD-25 to Banana/BNC power cable. Connect the Db-25 on the detector side, and on the power supply connect the ANA-24V and Dig-24V banana cables to the same voltage positive output, and set the output to 24V and max current of 3.5Amps. Connect the AGND and DGND to the negative output. You may want to connect the negative port to the ground port using banana cable. The BIAS voltage can either be floating or connected to the Keithley. For initial tests, set the Keithley to -1V and max current to 105uA. The final configuration should look like this

...