Some detectors have a stand-alone version of the DAQ process that can be run on the server the detector is connected to. This is the case for the Jungfrau, the Alvium and the Zyla for example. This is very helpful to decouple DAQ problem from detector setup issues.

The binaries can be found in /cds/group/pcds/dist/pds/current/build/pdsapp/bin/x86_64-rhel7-opt/. The detectors can be tested in free-run and triggered mode (the DAQ must be running for that). The stand-alone processes take similar argument as the DAQ process. It is a good idea to check the cnf and pass the same arguments.

...