Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Archon controls are on 2 screens currently. The "parent" screen (left) is accessed through the home screen (labeled "CCD"), and the "child" screen (right) is accessed by pressing "STA5200". Power, bias voltage, and heater controls are on the child screen. Under "Power", turn power and bias voltage on. Under "Heater", set the target temperature to -100C and turn the heater on. Here are the other heater settings, these should not be changed. 

...

Note: make sure to "Stop" acquisition when changing settings, and "Start" after to apply changes. (image highlighting Acquire button)

Settings to change1d2d
Readout ModeFVBImage
Bins1, 11, 1
Image Size 4800, 14800, 1200
Fixed Rate100Hz1Hz 
LineScan ModeEnableEnable

The first 3 parameters can be found under "Readout" and "Collect" in the archon child screen, and "Fixed Rate" is on the parent screen. 

Image Added  Image Added(images of archon screens, highlight fixed rate/binning/readout mode/what else?)




DAQ/AMI: 








NOTE: everything in the above screens is "expert mode". Users (and POCs for the most part) should not need to access all of this information. Controls group is working IOCs with interlocks to simplify CCD operation (add image of system and interlocks)







Running the CCD in the ASC