Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • ssh mec-laser
  • mecpython
  • import meclas
  • meclas.YFE.On() Note: Wait several minutes until the E-drives have turned on before proceeding to the next step. The E-drive turn on is confirmed after a series of chimes are heard. If not in the hutch, then the e-drive turn on can be confirmed by opening scope Lecroy A and changing the state of Long Pulse Laser Operation to 10Hz enable. The 10Hz trace will be visible when the E-drives have turned on (provided SHUTTER 4 on the LSM LSS is open).
    image-2023-7-7_12-52-40.pngImage Addedimage-2023-7-7_12-54-48.pngImage Added
  • meclas.LPL.SHG_opt()

...