- Make sure beam is off.
- logon to the rayonix machine from mfx-daq
ssh hsuser@con-ics-mfx-rayonix (password is "hsuser" if needed) - >killall procServ (this will kill the connection to the lcls daq)
- >capxure
- goto "Single Frame Collection" tab
- Set Trigger to "None" (if you don't do this, capxure may hang if you try to take frames). Set Exposure mode to "Dark".
- Change to your operating binning mode, and hit "Collect New Background"
- kill capxure (yes, it sometimes seg faults when you exit)
- cd ~/slac/, and do:
./startDaqInterface_newcraydl - On mfx-daq, set DAQ evr to run at 30Hz (if 4x4). Or use a sequencer if you're doing something fancy.
- Run
Connecting the Rayonix to the DAQ.
New from Kaz on
...