Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Motor Configuration

StageIOCPVCablingStatus
Newport UTS150PPV6ioc-mec-ppl-xps2MEC:PPL:MMN:13Standard Newport connection from mcn-mec-tch2.OK

Micos 6237-C002

Linear Stage, 40mm travel, HV, VSS-42 motor, 16:1 VPGL gear

Reference: https://www.pi-usa.us/en/products/precision-motorized-linear-stages

ioc-mec-usrppl-ims

MEC:USR:MMS:25

(formerly XTCS Rotation)

PR60 Patch Panel - DB9 Flange Connection "#4" - Stage

Note: Not connecting DB9 from Micos stage containing limit switch information.

PM: We measured the downward and upward motions in response to 5 mm requests. Downward is fine, but upward stops and starts. We recommend cleaning.

EG: We don't go to the limits when they are not reliable, we set soft limit 1mm before and after the limits to make sure we don't reach them.I'll be down in the hutch in the afternoon and can have a look at it.

Micronix VT-50-12316


https://micronixusa.com/product/download/7XReVO/universal-document/xVze6v/VT-50L_Datasheet_v1.5.pdf

VT-50L-12316

Motor Type
Stepper Motor
Limit Switch
Mechanical
Environmental Options
High Vacuum 10-6 mbar
Travel Range
100 mm
Encoder Option
Digital (RS-422)

ioc-mec-usrppl-ims

MEC:USR:MMS:20

(formerly XTCS Filter Wheel)

PR60 Patch Panel - DB15 Flange Connection "2" - Stage

PM: We have verified the cabling for the VT50 out to the cable that comes from the stage. Even when using a VT50 parameter set verified from CXI we saw no motion of the stage.Suggest mechanical inspection.



Detector Configuration

DetectorDAQ HostDAQ ConfigurationCablingStatus
Epix10kaQuad0daq-mec-pgp02-i 0 -P 0x10 -R 0 -D 0x1760Host card 0 - (via overhead cable tray) - back of PR60 - PR60 patch panel - Flange -  - DetectorOK
Epix10kaQuad1daq-mec-pgp02-i 1 -P 0x20 -R 0 -D 0x1760Host card 0 - (via overhead cable tray) - back of PR60 - PR60 patch panel - Flange -  - DetectorOK
Epix10kaQuad2daq-mec-pgp02-i 2 -P 0x11 -R 0 -D 0x1760Host card 1-  (via overhead cable tray - back of PR60 - PR60 patch panel - Flange  - DetectorOK
Epix10kaQuad3daq-mec-pgp02-i 3 -P 0x21 -R 0 -D 0x1760Host card 0 - (via overhead cable tray) - back of PR60 - PR60 patch panel - Flange - DetectorOK
epix100a_1_XTCSdaq-mec-cam01-i 1 -P 0x20 -G -T -R 0 -D 0x1770Host card 0 - R60 E33 RHS patch panel (connection #3) - (via overhead cable tray) back of PR60 - PR60 patch panel - Test Flange - Detector

Pending resolution of host FEZ interface issues, see:

Jira
serverSLAC National Accelerator Laboratory
serverId1b8dc293-975d-3f2d-b988-18fd9aec1546
keyECS-3745

Von Hamos Configuration

Components:

  • Remote (vncviewer) access to Windows PC
    • Requires Windows machine connected to ICS subnet.
  • Windows PC 
  • Cable from PC to Controller 
  • Controller 
  • Cable from Controller to flange 
  • Cable from flange to Crystals - in vacuum
  • Crystals - in vacuum