Info |
---|
This assumed Iota controller is set to external triggers and CYCLE else this does nothing. |
- Open the new colorful gas jet control panel from facethome: LI20 → Experiment → Gas Jet Controls...
- Set the rate by selecting the rate
- Change state to enabled to fire the jet. Disabled to stop.
- If the "Gas 1 & 2 Supply" indicator goes red then there is an EPS fault:
- Stop the jets
- Check there is no DPS issue
- Wait for the pressure to decrease
- Hit "RESET" to reset the EPS. The gas indicator should turn green
- More details can be found in the E-300 Oven panel
- If the Beamline valve indicators go red, then there is a beamline pressure problem.
- Stop the gas jets
- Ensure all the turbos are running - all turbo indicators are green, or see more details in DPS Full Schematic
- Wait for the pressures indicated to decrease and change back to green
- Open the Valve Intlks page and reset interlocks, then you can reopen the beamline valve from the DPS Full Schematic page if all valve interlocks are cleared
- If any turbo indicators turn red then call for help. More details can be found in the DPS Full Schematic
These are the old panel instructions:
- Go to the EVR panel: Go to facethome and LI20-Experimenter. From there, select the Triggers button (the screen shows items grouped by racks- these electronics are located in rack FKG20-24).
...