...
Info |
---|
Check all diodes at the beginning of the shift (ideally this should also be done a few days before the experiment to give you time to fix things that don't work), by plugging them into Lecroy and visual check (with bias-T). |
Install a fast diode and mirror (roughly at 45 degrees with respect to the beam) on the SC1 post_sample_x stage (CXI:SC1:MMS:02.RBV) downstream on the interaction region. There is a danger of driving this stage into the Jungfrau or other components in SC1, so it might be disabled. It is also good practice to disable the motor when not in use. Toggle Cloak
Cloak |
---|
The fast diode has a small area and is therefore can be difficult to align. Ideally, a Y stage (PP-30 piezo) is used to allow remote motion of the diode to align to the laser and x-rays. Due to the limited travel range of the long X-stage in SC1, the diode and mirror must be installed as far as reasonable towards the -X side, overhanging away from the carriage |
Toggle Cloak |
---|
Cloak |
---|
(-4 mm in position but this will soon be changed to 0 mm. If the "out" position is 92 mm, then -4 mm, if 96 mm, then 0 mm). |
Install a regular diode connected to the Acqiris on the sample chamber door and align the reflected reference laser beam on the mirror onto that diode. Toggle Cloak
...
Toggle Cloak |
---|
Cloak |
---|
Insert a nozzle rod with both a clear YAG and a frosted YAG. Toggle Cloak
...
Toggle Cloak |
---|
Cloak |
---|
Zoomed-in view: |
Rough timing
...
Toggle Cloak |
---|
Cloak |
---|
|
Toggle Cloak |
---|
Turn off Shutter the fs laser and turn on unshutter the X-rays. Move the diode around in x and y to maximize the X-ray signal (if the steps are very large then the laser may need to be repointedre-pointed) Toggle Cloak
Save Click Store on the Lecroy to save a trace of the X-rays on the scope Toggle Cloak
Turn Shutter X-rays off and turn on unshutter fs laser. Toggle Cloak
Toggle Cloak |
---|
Cloak |
---|
If needed, align the fs laser to the diode (usually not needed). Move laser delay time (LAS:R52B:EVR:31:TRIG0:TDES) to bring laser within 100 picoseconds of the X-rays using the scope |
...
. Alternatively, change the Target Time in the fs laser window. Align the fs laser trace so that the leading edges of the traces line up. You can center the diode signal on the Lecroy by changing the Delay. Try to match the amplitude of the X-ray and fs laser signals. Type the time below the ns Offset in the fs laser window in the ns Offset box, which will reset the Target Time to 0. |
Fine timing
Once the laser is roughly timed roughly, it is time to move onto the finer timing measurement using a cross-correlation measurement
...