Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Typing cooler=on engages the cooler's own PID control loop.  Since we rely on the sensor's PID loop for temp control, having two operating simultaneously might lead to instability and is best avoided.  If the cold head's PID mode is needed, settings to control it are in the CryoTel manual referenced above.   



Read Noise Tests