Confluence will be unusable 23-July-2024 at 06:00 due to a Crowd upgrade.
...
check the status of the data moving here: Data Mover Monitoring, this page can be seen one the main "pswww" page as well. Take out the nodes with the issue, assuming your problem is limited to a single node. If it's wider spread, it might warrant a call.
13_11:46:45:2015-Nov-13 11:46:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.75,-156.15,52.25,-860.95,-870.25,-774.25,33.05,49.85,49.85,276,273,275,36,0,0,1200,00:00:04
12_15:30:30:2015-Nov-12 15:30:30,[VERBOSE ],RxDetector::FrameAcquisitionFrameProcessor(83, 0x7f55d403bc20,0x7f55d42d2230) - starting
12_15:30:30:FrameReady: Frame #84 size=7372800 timestamp=59989 ms
13_11:38:45:2015-Nov-13 11:38:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.65,-109.25,-109.95,34.05,42.65,45.45,34.75,49.85,49.85,276,273,275,37,0,0,1200,20:50:31
13_11:39:45:2015-Nov-13 11:39:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.65,-109.25,-109.95,34.05,42.35,45.45,34.85,49.85,49.85,276,273,275,36,0,0,1200,20:51:31
13_11:40:45:2015-Nov-13 11:40:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.95,-79.75,-156.15,52.05,-860.85,-870.25,-774.15,34.75,49.85,49.85,276,273,275,37,0,0,1200,00:00:04
13_11:41:45:2015-Nov-13 11:41:45,[STATUS ],[STATUS_DATA],-79.95,-79.35,-79.85,-79.65,-156.15,52.05,-860.85,-870.25,-774.15,33.95,49.85,49.85,276,273,275,36,0,0,1200,00:00:04
To recover from that, you need to stop the DAQ process (ctrl+x in the telnet localhost 30099 terminal window) and open the Rayonix software capxure.
Reboot the detector controller (there is a button for that). Re-enable cooling. You might have to quite this process and start again to see the now reasonable temperatures. Once this looks fine, you can quit this software and restart the DAQ process: (ctrl+R in the telnet localhost 30099 terminal window)
...