TempA (C)TempB (C)Rate (Hz)Abs DNNoise (DN)Kaz NoiseData
-120.0-100.42001568.71.521.82/home/detopr/archon/data/20201105/1d_200Hz_120C

-100.41001569.61.581.93/home/detopr/archon/data/20201105/1d_100Hz_120C

-100.4501570.71.682.04/home/detopr/archon/data/20201105/1d_50Hz_120C
-115.0-96.42001569.01.541.83/home/detopr/archon/data/20201105/1d_200Hz_115C

-96.41001570.11.601.93/home/detopr/archon/data/20201105/1d_100Hz_115C

-96.4501571.91.772.16/home/detopr/archon/data/20201105/1d_50Hz_115C
-110.0-92.22001570.31.561.94/home/detopr/archon/data/20201105/1d_200Hz_110C

-92.21001571.91.682.22/home/detopr/archon/data/20201105/1d_100Hz_110C

-92.1501574.11.852.34/home/detopr/archon/data/20201105/1d_50Hz_110C
-105.0-87.62001572.41.561.93/home/detopr/archon/data/20201105/1d_200Hz_105C

-87.61001574.21.672.17/home/detopr/archon/data/20201105/1d_100Hz_105C

-87.6501576.91.952.74/home/detopr/archon/data/20201105/1d_50Hz_105C
-100.0-82.92001578.51.782.17/home/detopr/archon/data/20201105/1d_200Hz_100C

-82.91001581.52.032.48/home/detopr/archon/data/20201105/1d_100Hz_100C

-82.9501588.22.522.96/home/detopr/archon/data/20201105/1d_50Hz_100C
-95.0-78.02001577.52.112.21/home/detopr/archon/data/20201105/1d_200Hz_95C

-78.01001581.52.392.53/home/detopr/archon/data/20201105/1d_100Hz_95C

-78.0501587.72.813.01/home/detopr/archon/data/20201105/1d_50Hz_95C
-90.0-73.12001553.62.662.29/home/detopr/archon/data/20201105/1d_200Hz_90C

-73.11001556.72.862.62/home/detopr/archon/data/20201105/1d_100Hz_90C

-73.1501562.83.182.92/home/detopr/archon/data/20201105/1d_50Hz_90C
-85.0-68.62001502.82.772.30/home/detopr/archon/data/20201105/1d_200Hz_85C

-68.71001505.12.912.54/home/detopr/archon/data/20201105/1d_100Hz_85C

-68.7501510.03.222.58

/home/detopr/archon/data/20201105/1d_50Hz_85C

-80.0-64.72001433.02.342.24/home/detopr/archon/data/20201105/1d_200Hz_80C

-64.71001435.42.502.45/home/detopr/archon/data/20201105/1d_100Hz_80C

-64.7501441.52.832.90

/home/detopr/archon/data/20201105/1d_50Hz_80C

-75.0-60.92001372.11.932.36/home/detopr/archon/data/20201105/1d_200Hz_75C

-61.01001374.51.982.37/home/detopr/archon/data/20201105/1d_100Hz_75C

-61.0501382.22.402.71/home/detopr/archon/data/20201105/1d_50Hz_75C
0.02.22001258.78.177.29/home/detopr/archon/data/20201105/1d_200Hz_0C

2.21001366.318.59.88/home/detopr/archon/data/20201105/1d_100Hz_0C

2.1501614.647.513.51/home/detopr/archon/data/20201105/1d_50Hz_0C
10.011.42001500.412.712.8/home/detopr/archon/data/20201105/1d_200Hz_plus10C

11.41001931.021.916.8/home/detopr/archon/data/20201105/1d_100Hz_plus10C

11.4502895.441.924.3/home/detopr/archon/data/20201105/1d_50Hz_plus10C


Image Added


Surprising to me that the noise and leakage never really seemed to take off until well above -75C....expected to see it increase by x2 per 7K or some such, but I guess that's only around room T.  Will need to look at the data to understand how I'm confusing myself.

...