Method: Define noise stats box to include a block of pixels avoiding the non-pixel gaps in the display.  Since the block contains sequestial reads of the binned pixels, the standard deviation over the block is a proxy ~proxy for noise.  Record std. dev and absolute singal at 50 Hz, 100Hz, and 200Hz to measure leakage.  Gradually increase temp from -120C, pausing 15min at each temp to stabilize before recording.  Save 10 frames at each setting for later analysis.


TempA (C)TempB (C)Rate (Hz)Abs DNNoise (DN)Signal (DN)Data
-120.0-100.42001568.71.52
/home/detopr/archon/data/20201105/1d_200Hz_120C

-100.41001569.61.580.01/home/detopr/archon/data/20201105/1d_100Hz_120C

-100.4501570.71.68-0.03/home/detopr/archon/data/20201105/1d_50Hz_120C
-115.0-96.42001569.01.540.02/home/detopr/archon/data/20201105/1d_200Hz_115C

-96.41001570.11.60-0.02/home/detopr/archon/data/20201105/1d_100Hz_115C

-96.4501571.91.77-0.01/home/detopr/archon/data/20201105/1d_50Hz_115C
-110.0-92.22001570.31.560.05/home/detopr/archon/data/20201105/1d_200Hz_110C

-92.21001571.91.680.01/home/detopr/archon/data/20201105/1d_100Hz_110C

-92.1501574.11.850.00/home/detopr/archon/data/20201105/1d_50Hz_110C
-105.0-87.62001572.41.560.00/home/detopr/archon/data/20201105/1d_200Hz_105C

-87.61001574.21.670.02/home/detopr/archon/data/20201105/1d_100Hz_105C

-87.6501576.91.950.19/home/detopr/archon/data/20201105/1d_50Hz_105C
-100.0-82.92001578.51.780.02/home/detopr/archon/data/20201105/1d_200Hz_100C

-82.91001581.52.030.06/home/detopr/archon/data/20201105/1d_100Hz_100C

-82.9501588.22.520.03/home/detopr/archon/data/20201105/1d_50Hz_100C
-95.0-78.02001577.52.110.04/home/detopr/archon/data/20201105/1d_200Hz_95C

-78.01001581.52.390.00/home/detopr/archon/data/20201105/1d_100Hz_95C

-78.0501587.72.810.10/home/detopr/archon/data/20201105/1d_50Hz_95C
-90.0-73.12001553.62.660.07/home/detopr/archon/data/20201105/1d_200Hz_90C

-73.11001556.72.860.07/home/detopr/archon/data/20201105/1d_100Hz_90C

-73.1501562.83.180.11/home/detopr/archon/data/20201105/1d_50Hz_90C
-85.0-68.62001502.82.770.11/home/detopr/archon/data/20201105/1d_200Hz_85C

-68.71001505.12.910.04/home/detopr/archon/data/20201105/1d_100Hz_85C

-68.7501510.03.220.02

/home/detopr/archon/data/20201105/1d_50Hz_85C

-80.0-64.72001433.02.340.04/home/detopr/archon/data/20201105/1d_200Hz_80C

-64.71001435.42.500.04/home/detopr/archon/data/20201105/1d_100Hz_80C

-64.7501441.52.830.01

/home/detopr/archon/data/20201105/1d_50Hz_80C

-75.0-60.92001372.11.930.00/home/detopr/archon/data/20201105/1d_200Hz_75C

-61.01001374.51.980.02/home/detopr/archon/data/20201105/1d_100Hz_75C

-61.0501382.22.400.02/home/detopr/archon/data/20201105/1d_50Hz_75C
0.02.22001258.78.170.07/home/detopr/archon/data/20201105/1d_200Hz_0C

2.21001366.318.50.22/home/detopr/archon/data/20201105/1d_100Hz_0C

2.1501614.647.51.00/home/detopr/archon/data/20201105/1d_50Hz_0C
10.011.42001500.412.70.01/home/detopr/archon/data/20201105/1d_200Hz_plus10C

11.41001931.021.90.17/home/detopr/archon/data/20201105/1d_100Hz_plus10C

11.4502895.441.890.07/home/detopr/archon/data/20201105/1d_50Hz_plus10C

Surprising to me that the noise and leakage never really seemed to take off....expected to see it increase by x2 per 7K or some such, but I guess that only around room T.  Will need to look at the data to understand how I'm confusing myself.

...