2020-06-18 Calibration procedure for pedestals using charge injection

...

Reconstruction of data in analysis

<calibrated-data> = (<raw-data> - pedestals) * <gain-factor>

where pedestals and <gain-factor> are per-pixel map combined from 7 gain mode maps.

Calibration preocedure

Accumulate dark data and evaluate pedestals for five gain modes:

...

Assuming that each gain mode intensity sample is parametrized as
<intensity> = offset B + <charge-index> * <gain>G
each pixel has constants like

...

  • pedestal_AHL-L = pedestal_AHL-H  - offset_ahl
  • pedestal_AML-L = pedestal_AML-H - offset_aml

Image RemovedImage Removed 

 

Image AddedImage Added

Reconstruction of data in analysis

<calibrated-data> = (<raw-data> - pedestal) * <gain-factor>

 

Panel results

detdaq18 run 52

Panel 2(counting from 0) /reg/g/psdm/detector/gains/epix10k/panels/0000000001-0175233793-0553648150-0743498629-0017156647-0000000000-0000000000

Pedestals for fixed gain modes FH, FM, FL

Image AddedImage AddedImage Added

Pedestals for switching gain modes AHL-H, AML-M and derived AHL-L, AML-L

 

Image AddedImage AddedImage AddedImage Added

Offsets for switching gain modes AHL-H, AHL-L, AML-M, AML-L

Image AddedImage AddedImage AddedImage Added

Offsets difference  H(or M) - L for  AHL, AML

Image AddedImage Added

Gains for switching gain modes AHL-H, AHL-L, AML-M, AML-L

Image AddedImage AddedImage AddedImage Added

 where pedestal and <gain-factor> are per-pixel map combined from 7 gain mode maps.

Questions

  • is current procedure correct?
  • if not, what is correct procedure?
  • or what could we calibrate with charge injection?

...