Comments from Gabriel Blaj:

Panel

 

Metrology files:

2017-11-03-CSPAD-XPP-Camera3-Metrology.xlsx - WARNING: it has unusual numeration of quads (3,0,1,2), which is corrected in the text file

2017-11-03-CSPAD-XPP-Camera3-Metrology.txt

Quick quality check

Command: optical_metrology_check 2017-11-03-CSPAD-XPP-Camera3-Metrology.txt generates a few warnings:

Code Block
X-Y quality check for optical metrology measurements
  WARNING segm  8:  |-77| > 60.0
  WARNING segm  9:  |-82| > 60.0
  WARNING segm 19:  |64| > 60.0
  WARNING segm 29:  |72| > 60.0
Z quality check for optical metrology measurements 
  WARNING segm 12:  |341.0| > 100.0
  WARNING segm 23:  |-151.0| > 100.0
  WARNING segm 30:  |120.1| > 100.0

X-Y deviations are relatively small and are ignored.

Z deviations do not have any redundancy for obvious correction

Quality check in X-Y plane

...

(tick) Archive: /reg/g/psdm/detector/alignment/cspad/calib-xpp-2017-11-03-camera3/calib/CsPad::CalibV1/XppGon.0:Cspad.0/geometry/0-end.data