Back to CSPAD Alignment
Content
Optical measurements
Comments from Gabriel Blaj: CSPAD 2.3M (in air) Repair Oct 2017 Traveller.docx
![](/download/attachments/228306234/CSPAD%202.3M%20%28in%20air%29%20Repair%20Oct%202017.png?version=1&modificationDate=1510262117000&api=v2)
![](/download/attachments/228306234/XPP-Camera_3_20171103.png?version=1&modificationDate=1510262119000&api=v2)
Metrology files:
2017-11-03-CSPAD-XPP-Camera3-Metrology.xlsx - WARNING: it has unusual numeration of quads (3,0,1,2), which is corrected in the text file
2017-11-03-CSPAD-XPP-Camera3-Metrology.txt
Quick quality check
Command: optical_metrology_check 2017-11-03-CSPAD-XPP-Camera3-Metrology.txt
generates a few warnings:
X-Y quality check for optical metrology measurements
WARNING segm 8: |-77| > 60.0
WARNING segm 9: |-82| > 60.0
WARNING segm 19: |64| > 60.0
WARNING segm 29: |72| > 60.0
Z quality check for optical metrology measurements
WARNING segm 12: |341.0| > 100.0
WARNING segm 23: |-151.0| > 100.0
WARNING segm 30: |120.1| > 100.0
X-Y deviations are relatively small and are ignored.
Z deviations do not have any redundancy for obvious correction
Quality check in X-Y plane
2017-11-09 10:38:13 (info) CalibManager.GUIMetrology: Quality check in X-Y plane:
segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL)
Quad 0
segm: 0 20900 20901 -219 -218 43543 43578 73 108 0.28739 48302 48328 -26 -1 -35
segm: 1 20888 20934 -260 -214 43564 43566 104 106 0.31169 48316 48332 -16 -46 -2
segm: 2 -20918 -20925 98 91 43546 43556 45 55 0.12432 48319 48311 8 7 -10
segm: 3 -20923 -20917 -48 -42 43548 43549 -19 -18 -0.05921 48315 48310 5 -6 -1
segm: 4 -20904 -20909 143 138 -43541 -43556 -67 -82 0.18485 48313 48300 13 5 15
segm: 5 -20900 -20919 19 0 -43572 -43544 -2 26 0.01250 48301 48331 -30 19 -28
segm: 6 -20965 -20907 54 112 43552 43550 29 27 0.10919 48311 48332 -21 -58 2
segm: 7 -20897 -20916 39 20 43558 43531 22 -5 0.03882 48298 48308 -10 19 27
Quad 1
segm: 0 -21005 -21087 6 -76 43609 43602 -22 -29 -0.04599 48411 48426 -15 82 7
segm: 1 -20997 -20982 -96 -81 43618 43541 22 -55 -0.11635 48394 48347 47 -15 77
segm: 2 -20897 -20913 29 13 -43560 -43551 -14 -5 0.02762 48312 48313 -1 16 -9
segm: 3 -20910 -20925 -118 -133 -43563 -43561 67 69 -0.16507 48317 48330 -13 15 -2
segm: 4 20912 20911 73 72 -43541 -43547 32 26 -0.09540 48311 48299 12 1 6
segm: 5 20936 20920 94 78 -43555 -43548 37 44 -0.11314 48319 48318 1 16 -7
segm: 6 -20904 -20907 -119 -122 -43566 -43550 56 72 -0.15850 48309 48320 -11 3 -16
segm: 7 -20911 -20909 17 19 -43563 -43549 -2 12 0.02368 48302 48327 -25 -2 -14
Quad 2
segm: 0 -20907 -20902 -151 -146 -43542 -43606 116 52 -0.19526 48317 48340 -23 -5 64
segm: 1 -20911 -20895 -14 2 -43559 -43557 0 2 -0.00789 48315 48312 3 -16 -2
segm: 2 20903 20903 195 195 -43536 -43556 103 83 -0.25657 48304 48303 1 0 20
segm: 3 20903 20921 200 218 -43548 -43552 105 101 -0.27497 48308 48313 -5 -18 4
segm: 4 20906 20901 143 138 43540 43568 -73 -45 -0.18483 48318 48303 15 5 -28
segm: 5 20907 20909 216 218 43552 43544 -107 -115 -0.28550 48301 48313 -12 -2 8
segm: 6 20931 20906 -29 -54 -43545 -43529 -24 -8 0.05462 48298 48305 -7 25 -16
segm: 7 20925 20911 137 123 -43562 -43539 58 81 -0.17103 48307 48320 -13 14 -23
Quad 3
segm: 0 20921 20921 -98 -98 -43527 -43549 -26 -48 0.12897 48294 48312 -18 0 22
segm: 1 20932 20929 -74 -77 -43594 -43522 -90 -18 0.09931 48341 48310 31 3 -72
segm: 2 20896 20913 -125 -108 43554 43555 50 51 0.15325 48306 48316 -10 -17 -1
segm: 3 20910 20956 -68 -22 43560 43560 12 12 0.05919 48320 48337 -17 -46 0
segm: 4 -20923 -20912 -26 -15 43531 43537 -6 0 -0.02698 48304 48292 12 -11 -6
segm: 5 -20920 -20906 -106 -92 43547 43573 -47 -21 -0.13022 48332 48308 24 -14 -26
segm: 6 20900 20913 -111 -98 43550 43538 53 41 0.13750 48300 48305 -5 -13 12
segm: 7 20920 20920 41 41 43534 43565 -28 3 -0.05394 48320 48307 13 0 -31
Quality check in Z
2017-11-09 10:38:13 (info) CalibManager.GUIMetrology: Quality check in Z:
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um)
Quad 0
segm: 0 20900 43560 20900 43560 13 57 -16 28 35 6 -44 -44 35 6 0.09595 0.00789 43.977
segm: 1 20911 43565 20911 43565 27 -25 37 -15 1 11 52 52 1 11 0.00274 0.01447 -52.059
segm: 2 -20921 43551 43551 20921 -40 -3 -33 4 -21 -14 -37 -37 -14 -21 -0.01842 -0.05751 -37.000
segm: 3 -20920 43548 43548 20920 -92 -79 61 74 -85 67 -13 -13 67 -85 0.08815 -0.23280 -12.976
segm: 4 -20906 -43548 20906 43548 -16 -17 -42 -43 -16 -42 1 1 -16 -42 -0.04385 -0.05526 -0.981
segm: 5 -20909 -43558 20909 43558 -46 23 21 90 -11 55 -69 -69 -11 55 -0.03014 0.07235 68.993
segm: 6 -20936 43551 43551 20936 20 1 11 -8 10 1 19 19 1 10 0.00132 0.02737 18.998
segm: 7 -20906 43544 43544 20906 31 -30 47 -14 0 16 61 61 16 0 0.02105 0.00000 60.981
Quad 1
segm: 0 -21046 43605 43605 21046 29 23 2 -4 26 -1 6 6 -1 26 -0.00131 0.07078 -6.114
segm: 1 -20989 43579 43579 20989 27 -36 -17 -80 -4 -48 63 63 -48 -4 -0.06311 -0.01092 -63.122
segm: 2 -20905 -43555 20905 43555 -117 -27 -34 56 -72 11 -90 -90 -72 11 -0.19733 0.01447 -90.032
segm: 3 -20917 -43562 20917 43562 -59 7 -51 15 -26 -18 -66 -66 -26 -18 -0.07122 -0.02367 -65.996
segm: 4 20911 -43544 43544 20911 343 2 401 60 172 230 341 341 230 172 0.30263 0.47127 -341.041
segm: 5 20928 -43551 43551 20928 -35 -59 6 -18 -47 -6 24 24 -6 -47 -0.00789 -0.12867 -24.044
segm: 6 -20905 -43558 20905 43558 -40 23 3 66 -8 34 -63 -63 -8 34 -0.02193 0.04472 -63.021
segm: 7 -20910 -43556 20910 43556 13 7 19 13 10 16 6 6 10 16 0.02740 0.02105 5.995
Quad 2
segm: 0 -20904 -43574 20904 43574 -63 20 -100 -17 -21 -58 -83 -83 -21 -58 -0.05756 -0.07626 83.009
segm: 1 -20903 -43558 20903 43558 13 -10 -8 -31 1 -19 23 23 1 -19 0.00274 -0.02499 -22.991
segm: 2 20903 -43546 43546 20903 -2 -26 -16 -40 -14 -28 24 24 -28 -14 -0.03684 -0.03837 23.982
segm: 3 20912 -43550 43550 20912 -14 -48 12 -22 -31 -5 34 34 -5 -31 -0.00658 -0.08494 33.957
segm: 4 20903 43554 20903 43554 2 -18 33 13 -8 23 20 20 -8 23 -0.02193 0.03026 -20.026
segm: 5 20908 43548 20908 43548 -79 72 -86 65 -3 -10 -151 -151 -3 -10 -0.00822 -0.01316 150.976
segm: 6 20918 -43537 43537 20918 5 26 15 36 15 25 -21 -21 25 15 0.03290 0.04109 -21.044
segm: 7 20918 -43550 43550 20918 28 24 -17 -21 26 -19 4 4 -19 26 -0.02500 0.07122 3.995
Quad 3
segm: 0 20921 -43538 43538 20921 1 37 -53 -17 19 -35 -36 -36 -35 19 -0.04606 0.05203 36.009
segm: 1 20930 -43558 43558 20930 -7 20 -117 -90 6 -103 -27 -27 -103 6 -0.13548 0.01642 26.850
segm: 2 20904 43554 20904 43554 117 -3 103 -17 57 43 120 120 57 43 0.15623 0.05657 119.997
segm: 3 20933 43560 20933 43560 38 35 80 77 36 78 3 3 36 78 0.09854 0.10260 3.077
segm: 4 -20917 43534 43534 20917 -63 -75 56 44 -69 50 12 12 50 -69 0.06581 -0.18900 -12.047
segm: 5 -20913 43560 43560 20913 -48 -36 18 30 -42 24 -12 -12 24 -42 0.03157 -0.11507 11.965
segm: 6 20906 43544 20906 43544 8 -12 11 -9 -2 1 20 20 -2 1 -0.00548 0.00132 19.995
segm: 7 20920 43549 20920 43549 -54 33 -52 35 -10 -8 -87 -87 -10 -8 -0.02739 -0.01053 -86.975
Constants
Archive: /reg/g/psdm/detector/alignment/cspad/calib-xpp-2017-11-03-camera3/calib/CsPad::CalibV1/XppGon.0:Cspad.0/geometry/0-end.data
- exp=mfx11116:run=624-626
- MfxEndstation.0:Cspad.0 and MfxEndstation.0:Jungfrau.0
IP 0 CSPAD:V2 0 -77500 120000 1000000 0.0 0.0 0.0 0.00000 0.00000 0.00000
IP:V1 0 JFCAMERA:V1 0 -90000 -3000 100000 90 0 0 0.0 0.0 0.0
![](/download/attachments/228306234/img-geo-cspad-mfx.png?version=1&modificationDate=1512092975000&api=v2)
![](/download/attachments/228306234/img-geo-jf-mfx.png?version=1&modificationDate=1512092977000&api=v2)
References