CU ACD have basically two calibration parameters for each channel (PMT), 6 in total - MIP peak position and pedestal (both - in ADC bins). Calibration parameters (pedestals, MIP peak positions and tiles light yield) for PS obtained in the 5 GeV pions runs on July 29, are given in the attachment. These parameters are preliminary and are subject for the revision. Please use them with caution.

 Calibration parameters for SPS runs, obtained in 150 GeV proton runs 700001865-700001871 are given in table below. Explanation for the Tile 3 is given in the attachment "ACD_Tile_3_calibration.doc"

Tile Physical #

Tile Recon Id

GAFE

Pedestal

MIP peak, pedestal removed

0

130

13

140

1396

1

120

11

141

1928

2

110

9

167

908

3A

100A

15

1033

Use combination 1.3*PMT_A+PMT_B
with MIP peak at 1623

3B

100B

17

567

 

4

0

 

212

492

 and on 150 GeV proton beam on SPS, are given in the attachment\.

Light Yield (Number of PhotoElectrons)

...