or the same in um

Code Block
 21757   21769   33464   10769   68489   68561   77637   54810  
 21788   21783   33479   10689   68469   68489   77863   54430  
 21678   21820   33054   10637   68310   68254   77976   54617  
 21851   21796   33444   11150   68801   68743   77582   54648  

 33110   10457   68275   68299   56732   79628   21754   21558  
 33319   10429   68135   68182   56664   79453   21668   21628  
 33141   10369   68476   68506   57121   79700   21858   21839  
 33030   10402   68338   68378   56684   79429   21605   21540  

    51      18     -28      18      71     -20     -15     -54  
    39     -18      -7       3      41     -14      -4     -20  
     1     -40     104     -24       9     -82      23       6  
    -5      -8      27      20       2     -23      23      13

 

(tick) tilt/0-end.data (degree)

No Format
 0.04474   0.01053  -0.01645  -0.02828  -0.05128  -0.03552  -0.33658  -0.06315  
 0.03750  -0.06382  -0.06119  -0.00197  -0.07697   0.01381   0.15327   0.05528  
-0.03948  -0.23876  -0.10659  -0.06711   0.16514   0.11842   0.09802   0.19144  
-0.10922  -0.13684   0.13617  -0.17232  -0.07560  -0.00263   0.27292   0.24731 

...

No Format
0    0  820  820
0  820  820    0
0    0    0    0

 

 

 

 

 

 

2014-03-05 CSPAD CxiDs1 alignment

Data for alignment

In order to get cspad geometry parameters for experiment cxib6714 Jason Koglin suggested to use cxi83714 runs 17,18, 127, 128, 130-132, 135, 136.

CSPAD array with subtracted pedestal and threshold 4*rms(per pixel) is averaged over 5000 events of run 136.

Parameters from latest optical measurement

center and tilt are used from /reg/neh/home1/dubrovin/LCLS/CSPadAlignment-v01/calib-cxi-ds1-2013-12-20

 

Tuned calibration parameters

offset_corr/0-end.data :

Code Block
  0   0    4   4
  0   4    1   1
  0   0    0   0

 

marg_gap_shift/0-end.data :

Code Block
15   40   4  40
15   40   4  40
 0    0   0   0

Results

2014-03-05 Alignment of Quads

Data for alignment

In order to get cspad geometry parameters for experiment cxib6714 Jason Koglin suggested to use cxi83714 runs 17,18, 127, 128, 130-132, 135, 136.

CSPAD array with subtracted pedestal and threshold 4*rms(per pixel) is averaged over 5000 events of run 136.

Image RemovedImage RemovedImage AddedImage Added

Parameters from latest optical measurement

center and tilt are used from /reg/neh/home1/dubrovin/LCLS/CSPadAlignment-v01/calib-cxi-ds1-2013-12-20

 

Tuned calibration parameters

offset_corr/0-end.data :

...

Code Block
  0   0    4   4
  0   4    1   1
  0   0    0   0

 

marg_gap_shift/0-end.data :

...

Code Block
15   40   4  40
15   40   4  40
 0    0   0   0

...

Results

Image RemovedImage RemovedImage Removed

 

 

 

 

 Image AddedImage AddedImage Added

 

2014-03-26 quads alignment with semi-automatic algorithm for ring center finding

...