Cherenkov counters settings

Cherenkov Settings

Electron runs:

- Discriminator threshold:  C1=140 mV, C2=150mV

- Pressure settings- For electron runs:

 Energy

 C1 pressure

C2 pressure

 0.5 GeV

 0.9 bars

 0.95 bars

 1

 0.9

 0.95

 2

 0.8

 0.95

 3

 0.8

 0.95

 4

 0.5

 0.6

 5

 0.3

 0.3

 6

 0.22

 0.22

 7

 0.18

 0.18

 

 

 

 

 

 

Proton runs:

- Discriminator threshold: C1= 30 mV,  C2 = 30 mV  

- Pressure settings For proton runs:  

 Energy

 C1 pressure

 C2 pressure

6 - 10 GeV

 2 bars

 2 bars

 

Instructions to set Cherenkov counters pressure

Note: at least once a day the cherenkov counters need to be completely evacuated and refilled. This operation takes about half an hour.

The Cherenkov rack is divided in two parts: the upper part for C1, the lower part for C2. All the operations must be performed on both Cs.

 How to empty Cherenkov counters:

...

- Go back to the general Cherenkov rack and close the main T9 cherenkov lever (horizontal position) 

 Thresholds