The following is a digest of the T9 runs. Get the full list or the selected run from the pipeline or elogbook. (WORK IN PROGRESS).

In red color : Not to be used, due to mistake, or for specialist. 

 Electron runs

 Runs

 Date

particle

energy

CU configuration

CU position

AD

notes

 1201

 11/8

e- 

 2.5

 BT1

 0°

 

 av rate 34 HZ, CU instant rate 500 hz. S0 counts 21k per spîll

 1202

 

 e-

 2.5

 BT1

 

 

 3 phasing errors for 30k events. run stopped due to beam off

1205


e-

1.




S0:14 k count per spill
HW trg: 700 count per spill
DAQ trg: 300 count per spill
beam divergence: 6.3mrad on XZ, 5mrad on YZ

 1218-1219

11/08 
21h43 

 e-

 1.

 

 30° (x,y,z)=(200,13.2,-48)

no 

 Accidentally started runs in synchronized Run Control. These were automatically aborted by the program.

 1220

 22h06

 e-

1. 

BT1 

idem 

 No

 100K, S0 new trigger, S4 off

 1221

 22h20

 e-

 1.

BT1 

Idem 

 YES

 Refrence point before changing momentum

 

 

 

 

 

 

 

 

...

 Runs

 Date

particle

energy

IMagnet 

CU configuration

CU position

AD

notes

1176 

 

tagged g

3GeV

 600

BT1

0 degrees, center of tower 2

yes 

events in twr3?

 1177-1180

 

tagged g

e beam at 2.5GeV

 600

BT1

0 degrees, center of tower 2

yes

magnet ON, 25K evts per run 

 1206

 

 tagged gamma

e-  1GeV

 240

 

  idem

 

 S0:14k count per spill
HW trg: 100 count per spill
DAQ trg: 100 count per spill
Lot of 0 cluster in chamber 2 and 3, we need to tune the bending magnet and/or improve the veto to select a small spot.

 1207

 11/8

 tagged gamma

 

  240

 

  idem

 

 idem with S1 in the trigger
this run is to see if the zero that we see is a trigger or magner problem: it seem that it's a trigger issue: now the fraction of zero seems similar to what we saw in the other physics runs. 

 1208

11/8 11h34

 tagged gamma

 e-
1 GeV

 240

 

  idem

yes

  New trigger S3 in veto  - 25k

 1209

 11/8
13h29 

 tagged gamma

 e-
1GeV

 240

 

 idem

 yes and no

around 30k : SYNCHRONIZATION LOST !
stopping the run, first 30000 events should well sycnhronized...  

1210 

 11/8
16h34 

 tagged gamma

 e-
1GeV 

 240

 

 idem

 yes

 Synchronized Tagged gamma with 1GeV electron beam, going for 70k to get all the statistic : 100Kevents
(+25 k from 1208 this cpunt is not taken into account in the eloogbook)

 1211-1213

 11/8

  Tagged

 1 GeV

180-240

  BT1

 30° (x,y,z)=(200,13.2,-48)

 

 various Magnet currents for setting up

 1214

 

 Tagged

 e 1 GeV

 240

   BT1

 30° (x,y,z)=(200,13.2,-48)

 

 130k

 1215

 

 Tagged

 e 1GeV

 240

   BT1

 idem

 

 25k

 1216

 

 Tagged

 e 1 GeV

 240

   BT1

 idem

 

 25K

 1217

 11/8
21h22 

 Tagged

 e 1 GeV

 240

   BT1

 idem

 

 30k
No spill during 4 minute

 1232

 12/08

 Tagged

 e 2.5 GeV

 600

  BT1

 30° (x,y,z)=(200,13.2,-48

 

 Not to be used , S4 not in the Trigger

 1233-1237

12/8 
7:21-9:36 

 Tagged

 e 2.5

 600

   BT1

 idem

 

 

Full-brem Photon runs

 Runs

 Date

particle

energy

 I(Magnet)

CU configuration

CU position

AD

notes

1183-1190

 8/8
9/8 

e+g

2.5

 600

BT1

0 degrees, center of tower 2

yes 

overall 2M evts  - (false report of 300 A for 1185 in elogBook)

1222

 11/8
22:50 


  2.5 GeV

 0


30° (x,y,z)=(200,13.2,-48

yes


 1223-1231

12/8 

e+g 

2.5 GeV 

 600

   BT1

idem 

 

 rate stable after adjustement in run 1223
2.5 Mevents 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

...