IssueDescriptionDate AddedAdded byActions TakenStatus
FEB chillerDisconnected after found dirty5/6/2019tknReplaced, new chiller plumbed in and running.  Serial control is up but needs testing.unresolved
Wiener LV cardCard won't power - slot 4. On initial look, slots 4+ in crate had no power5/8/2019tknSwapped in spare card located in slot 10. See logbook entryresolved (OM)
FEBsNot all working5/8/2019tknSwapped L6B -> L4B (perfect) and put partially repaired FEB3 (three bad adc channels) in L6B. Can consider further replacement if perfect spares are achieved. (TKN)unresolved
AlignmentAlignment of L1-3 relative to wires is unknown5/19/2019tknSurvey/adjustment was performed by JLab alignment group to set box in same position as before relative to beamline. Motor positions were determined to set horizontal wires at beam height to create a reference point in lever sweep and vertical positions of lever ends were recorded at those positions. Positions of silicon relative to wires are still unknown and may require further outputs from the detector model as well as L0-3 survey data.unresolved
Cable LabelsLabel cabels to SVT/FEB chillers5/8/2019mrsolt unresolved
Chiller Serial CableRedo spliced connections of the serial cable to SVT/FEB chiller5/8/2019mrsolt unresolved
SVT Chiller WebcamSetup the SVT chiller webcam to monitor fluid levels5/8/2019mrsolt unresolved
Checked motor decodersNeed to understand if motor decoders need to be replaced5/9/2019omoreno unresolved
Check motor controllerVerify that motor moves the expected distance when moved from the GUI5/9/2019omoreno unresolved
Re-set motor home and operation positionNew layer 0 modules requires resetting motor positions.5/9/2019omoreno unresolved
Add SVT variables to EPICSBefore, values were retrieved by parsing the XML.  Now, rogue will provide us
the variables directly.
5/9/2019omoreno unresolved
Add RTD to SVT Chiller outletWe have a broken RTD on the SVT chiller outlet, and we should replace it5/9/2019mrsoltRTD removed and placed on table for survey. Need to have returned to SLAC. Need to order new RTD to get reworked. (TKN)unresolved