Issue mfxc00318 runs 10 and 13

Panel 6 sick asic  - long standing issue

detdaq18 r92 offset_AHL and AML

mfxc00118 r80 offset_AHL and AML

 

Charge injection re-calibration using mfxc00118 runs 80

Charge injection run 80
epix10ka_test_calibcycles exp=mfxc00118:run=80:smd --detname MfxEndstation.0:Epix10ka2M.0

Dark runs 82,86
epix10ka_test_calibcycles exp=mfxc00118:run=86:smd --detname MfxEndstation.0:Epix10ka2M.0

rm panels/*/*/*mfxc00118_r0080*

bsub -o 2020-08-14-r80-offset-log00 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i0 
bsub -o 2020-08-14-r80-offset-log01 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i1 
bsub -o 2020-08-14-r80-offset-log02 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i2 
bsub -o 2020-08-14-r80-offset-log03 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i3 


in ~3 hours 
bsub -o 2020-08-14-r80-offset-log04 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i4 
bsub -o 2020-08-14-r80-offset-log05 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i5 
bsub -o 2020-08-14-r80-offset-log06 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i6 
bsub -o 2020-08-14-r80-offset-log07 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i7 


in ~3 hours 
bsub -o 2020-08-14-r80-offset-log08 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i8 
bsub -o 2020-08-14-r80-offset-log09 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i9 
bsub -o 2020-08-14-r80-offset-log10 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i10
bsub -o 2020-08-14-r80-offset-log11 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i11


in ~3 hours 
bsub -o 2020-08-14-r80-offset-log12 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i12
bsub -o 2020-08-14-r80-offset-log13 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i13
bsub -o 2020-08-14-r80-offset-log14 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i14
bsub -o 2020-08-14-r80-offset-log15 -q psanaq epix10ka_offset_calibration -e mfxc00118 -d MfxEndstation.0:Epix10ka2M.0 -r80 -p -i15
 
 386 2020-08-14-r80-offset-log00
 386 2020-08-14-r80-offset-log01
 386 2020-08-14-r80-offset-log02
 386 2020-08-14-r80-offset-log03
 386 2020-08-14-r80-offset-log04
 386 2020-08-14-r80-offset-log05
5250 2020-08-14-r80-offset-log06
 386 2020-08-14-r80-offset-log07
 386 2020-08-14-r80-offset-log08
 386 2020-08-14-r80-offset-log09
 386 2020-08-14-r80-offset-log10
 386 2020-08-14-r80-offset-log11
 386 2020-08-14-r80-offset-log12
 386 2020-08-14-r80-offset-log13
 386 2020-08-14-r80-offset-log14
 386 2020-08-14-r80-offset-log15


In 2020-08-14-r80-offset-log06:

[W] L0249: pulser saw edges are not found, skip processing for ix41-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix42-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix43-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix44-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix45-iy24:
[W] L0249: pulser saw edges are not found, skip processing for ix46-iy24:

/reg/g/psdm/detector/gains/epix10k/panels/0000000001-0175696641-1056964630-1010951301-0019435010-0000000000-0000000000/offset/epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AML.dat

Replace sick-asic with 0-s by mean values for 3 healthy-asics in
epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AHL.dat
epix10ka_0032_20200808074716_mfxc00118_r0080_offset_AML.dat

mfxc00118 80 offsets AHL, AML

mfxc00118 80 pedestals AHL-H, AML-M, AHL-L, AML-L

 

mfxc00118 80 zero-offsets in sick asic manually corrected to mean value for 3 other asics for AHL and AML

Pedestal re-calibration for mfxc00118 run 86

pedestals AHL-H, AML-M, AHL-L, AML-L

 

Pedestals after run mfxc00118 run 80 and 86

epix10ka_offset_calibration for run 80

epix10ka_pedestals_calibration for run 86

Fixed pedestals for epix10ka_pedestals_calibration for run 86

pedestals with sick and fixed ASIC on panel6

to fix it one has to substitute bad ASIC constants for charge injection run for offset and gainci for AHL-H, AML-M, AHL-L, AML-L

All gains - Gabriel's constants and fixed panel 6 charge-injection gains from run 80

Comments on issue

New Gabriel's formulae was implemented in epix10ka_offset_calibration and not implemented in epix10ka_pedestals_calibration. Low pedestal PL was not evaluated consistently for two switching gain modes AHL and AML in the pedestal calibration script.

 

 

  • No labels