Content
Issue with non-uniform pulser intensity increment between events
Calib-cycle for AML-M
epix10ka_0026_20210510164634_mfxc00118_r0372-MfxEndstation-0-Epix10ka2M-0-seg00-cc008-AML-M-img-charge.png
Issue was seen earlier, recently reported by Philip due to poor fit Chi2.
epix10ka_0026_20210510164634_mfxc00118_r0372-MfxEndstation-0-Epix10ka2M-0-seg00-cc008-AML-M-fit-r014-c353-ibr02-ibc50.png
epix10ka_0026_20210510164634_mfxc00118_r0372-MfxEndstation-0-Epix10ka2M-0-seg00-cc008-AML-M-dat-r014-c353-ibr02-ibc50.png
Issues between events/records
- 694-695
- 2919-2920
- nevt - counter for any det.raw()
- nrec - counter for det.raw() is not None
- fid = evt.get(EventId).fiducials() normal increment between events is 3
nevt, nrec, fid: 0691 0691 102969
nevt, nrec, fid: 0692 0692 102972
nevt, nrec, fid: 0693 0693 102975
nevt, nrec, fid: 0694 0694 102978
nevt, nrec, fid: 0695 0695 103035
nevt, nrec, fid: 0696 0696 103038
nevt, nrec, fid: 0697 0697 103041
nevt, nrec, fid: 0698 0698 103044
nevt, nrec, fid: 2916 2916 109698
nevt, nrec, fid: 2917 2917 109701
nevt, nrec, fid: 2918 2918 109704
nevt, nrec, fid: 2919 2919 109707
nevt, nrec, fid: 2920 2920 109764
nevt, nrec, fid: 2921 2921 109767
nevt, nrec, fid: 2922 2922 109770
nevt, nrec, fid: 2923 2923 109773
Calib-cycle for AHL-H
Normal behavior
epix10ka_0026_20210510164634_mfxc00118_r0372-MfxEndstation-0-Epix10ka2M-0-seg00-cc057-AHL-H-dat-r014-c353-ibr02-ibc50.png
epix10ka_0026_20210510164634_mfxc00118_r0372-MfxEndstation-0-Epix10ka2M-0-seg00-cc057-AHL-H-fit-r014-c353-ibr02-ibc50.png