- Created by Mikhail Dubrovin, last modified on Feb 27, 2023
Back to CSPAD Alignment
Content
Optical measurements
Comments from Chris Kenney:
It is in the standard ePix 2M format,
except with each sensor "split" in to two halves.
Origin is outermost corner of Quad 3
Top of detector - where the coolant lines are - is at the top
Metrology files:
- 2023-02-22-CSPAD-CXI-SC3-Metrology.xlsx
- 2023-02-22-CSPAD-CXI-SC3-Metrology-orig.txt
- 2023-02-22-CSPAD-CXI-SC3-Metrology-corr.txt
- 2023-02-22-CSPAD-CXI-SC3-geometry-v2.txt
Quality check
> optical_metrology_check -i 2023-02-22-CSPAD-CXI-SC3-Metrology.txt
Quality check in X-Y plane
Quality check in X-Y plane Expand source
X-Y quality check for optical metrology measurements ---------------------------------------------------------------------------------------------------------------------------- segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) ---------------------------------------------------------------------------------------------------------------------------- segm: 0 20902 20906 -187 -183 43537 43539 85 87 -0.24346 48294 48299 -5 -4 -2 segm: 1 20904 20905 -136 -135 43541 43541 61 61 -0.17830 48295 48303 -7 -1 0 segm: 2 20905 20903 125 123 43539 43536 -60 -63 -0.16318 48294 48297 -3 2 3 segm: 3 20913 20903 126 116 43541 43536 -54 -59 -0.15923 48300 48297 2 10 5 segm: 4 20904 20902 3 1 43544 43544 0 0 0.00263 48302 48300 1 2 0 segm: 5 20908 21003 -386 -291 43542 43537 144 139 -0.44544 48302 48340 -38 -95 5 <<<< segm: 6 20906 20899 -24 -31 43538 43536 16 14 0.03619 48296 48293 3 7 2 segm: 7 20903 20900 111 108 43543 43536 -53 -60 -0.14410 48293 48300 -7 3 7 segm: 8 20910 20909 -89 -90 43543 43541 45 43 -0.11777 48303 48301 1 1 2 segm: 9 20908 20907 -19 -20 43550 43552 3 5 -0.02565 48304 48314 -9 1 -2 segm:10 20903 20907 0 4 43537 43536 2 1 -0.00263 48298 48293 4 -4 1 segm:11 20914 20912 76 74 43548 43545 -35 -38 -0.09868 48307 48308 0 2 3 segm:12 20905 20911 -144 -138 43551 43549 60 58 -0.18550 48301 48316 -15 -6 2 segm:13 20899 20894 -57 -62 43545 43544 35 34 -0.07829 48304 48293 10 5 1 segm:14 20876 20901 93 118 43540 43536 -61 -65 -0.13884 48278 48300 -22 -25 4 segm:15 20909 20903 -48 -54 43539 43538 27 26 0.06711 48299 48295 3 6 1 segm:16 20905 20903 87 85 43546 43545 -40 -41 0.11316 48303 48302 1 2 1 segm:17 20903 20906 -31 -28 43481 43486 6 11 -0.03887 48242 48252 -10 -3 -5 segm:18 20909 20910 90 91 43538 43541 -49 -46 -0.11909 48296 48303 -7 -1 -3 segm:19 20911 20924 -118 -105 43542 43547 46 51 0.14671 48303 48313 -9 -13 -5 segm:20 20907 20910 29 32 43550 43544 -9 -15 0.04013 48308 48304 4 -3 6 segm:21 20906 20909 -33 -30 43544 43545 13 14 -0.04145 48302 48305 -2 -3 -1 segm:22 20908 20908 2 2 43538 43541 -4 -1 -0.00263 48298 48300 -2 0 -3 segm:23 20912 20910 -110 -112 43542 43541 54 53 0.14606 48302 48302 0 2 1 segm:24 20912 20909 -36 -39 43545 43551 19 25 -0.04934 48311 48304 7 3 -6 segm:25 20905 20906 -108 -107 43545 43545 52 52 -0.14145 48303 48303 0 -1 0 segm:26 20890 20901 -215 -204 43531 43530 99 98 0.27575 48284 48288 -3 -11 1 segm:27 20910 20905 -253 -258 43539 43536 124 121 0.33624 48298 48298 0 5 3 segm:28 20905 20904 94 93 43540 43536 -43 -47 0.12305 48296 48296 0 1 4 segm:29 20897 20910 3 16 43542 43542 -1 -1 0.01250 48302 48296 6 -13 0 segm:30 20916 20908 -285 -293 43540 43537 143 140 0.38031 48303 48298 4 8 3 segm:31 20907 20911 -92 -88 43736 43538 46 -152 0.11817 48301 48474 -173 -4 198 <<<< ----------------------------------------------------------------------------------------------------------------------------
Deviations
Deviations
WARNING segm 5: |-95| > 60.0 WARNING segm 31: |-173| > 60.0 WARNING segm 31: |198| > 60.0
Quality check in Z
Quality check in Z Expand source
Z quality check for optical metrology measurements ----------------------------------------------------------------------------------------------------------------------------------------- segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) ----------------------------------------------------------------------------------------------------------------------------------------- segm: 0 20904 43538 43538 20904 -40 -42 74 72 -41 73 2 2 73 -41 0.09607 -0.11238 -1.996 segm: 1 20904 43541 43541 20904 -30 -29 109 110 -29 109 -1 -1 109 -29 0.14409 -0.08085 1.001 segm: 2 20904 43537 20904 43537 -2 -2 -89 -89 -2 -89 0 0 -2 -89 -0.00548 -0.11712 0.006 segm: 3 20908 43538 20908 43538 -1 10 -60 -49 4 -54 -11 -11 4 -54 0.01233 -0.07172 -10.993 segm: 4 20903 43544 43544 20903 -60 -75 27 12 -67 19 15 15 19 -67 0.02566 -0.18502 -15.006 segm: 5 20955 43539 43539 20955 -21 -33 -14 -26 -27 -20 12 12 -20 -27 -0.02632 -0.07382 -11.906 segm: 6 20902 43537 20902 43537 -10 -26 -158 -174 -18 -166 16 16 -18 -166 -0.04934 -0.21846 16.010 segm: 7 20901 43539 20901 43539 13 -21 -86 -120 -4 -103 34 34 -4 -103 -0.01096 -0.13554 34.012 segm: 8 20909 43542 43542 20909 -16 -20 -1 -5 -18 -3 4 4 -3 -18 -0.00395 -0.04932 -4.001 segm: 9 20907 43551 43551 20907 -25 -10 -9 6 -17 -1 -15 -15 -1 -17 -0.00197 -0.04796 14.999 segm:10 20905 43536 20905 43536 -36 -38 -220 -222 -37 -221 2 2 -37 -221 -0.10141 -0.29084 1.998 segm:11 20913 43546 20913 43546 -37 -26 -185 -174 -31 -179 -11 -11 -31 -179 -0.08630 -0.23617 -10.984 segm:12 20908 43550 43550 20908 -56 -54 -22 -20 -55 -21 -2 -2 -21 -55 -0.02763 -0.15072 2.015 segm:13 20896 43544 43544 20896 -164 -71 5 98 -117 51 -93 -93 51 -117 0.06776 -0.32217 92.958 segm:14 20888 43538 20888 43538 -9 9 -44 -26 0 -35 -18 -18 0 -35 0.00000 -0.04606 -18.007 segm:15 20906 43538 20906 43538 -16 -13 -20 -17 -14 -18 -3 -3 -14 -18 -0.03974 -0.02435 -2.995 segm:16 20904 43545 43545 20904 -4 -9 3 -2 -6 0 5 5 0 -6 0.00066 -0.01782 -5.000 segm:17 20904 43483 43483 20904 -9 -31 22 0 -20 11 22 22 11 -20 0.01449 -0.05482 -22.001 segm:18 20909 43539 20909 43539 23 43 -78 -58 33 -68 -20 -20 33 -68 0.09043 -0.08948 -20.004 segm:19 20917 43544 20917 43544 11 13 -149 -147 12 -148 -2 -2 12 -148 0.03287 -0.19474 -2.010 segm:20 20908 43547 43547 20908 -22 -32 29 19 -27 24 10 10 24 -27 0.03158 -0.07399 -9.993 segm:21 20907 43544 43544 20907 -32 -35 52 49 -33 50 3 3 50 -33 0.06645 -0.09180 -2.997 segm:22 20908 43539 20908 43539 9 12 -24 -21 10 -22 -3 -3 10 -22 0.02877 -0.02961 -3.002 segm:23 20911 43541 20911 43541 -49 -40 -73 -64 -44 -68 -9 -9 -44 -68 -0.12193 -0.09014 -8.994 segm:24 20910 43548 43548 20910 -5 -20 -28 -43 -12 -35 15 15 -35 -12 -0.04671 -0.03425 -14.997 segm:25 20905 43545 43545 20905 -26 63 -167 -78 18 -122 -89 -89 -122 18 -0.16118 0.05070 89.001 segm:26 20895 43530 20895 43530 41 42 -90 -89 41 -89 -1 -1 41 -89 0.11379 -0.11780 -0.976 segm:27 20907 43537 20907 43537 20 34 3 17 27 10 -14 -14 27 10 0.07399 0.01316 -14.005 segm:28 20904 43538 43538 20904 -22 -30 40 32 -26 36 8 8 36 -26 0.04738 -0.07126 -7.997 segm:29 20903 43542 43542 20903 3 -1 81 77 1 79 4 4 79 1 0.10395 0.00274 -4.002 segm:30 20912 43538 20912 43538 8 15 2 9 11 5 -7 -7 11 5 0.03151 0.00724 -7.003 segm:31 20909 43637 20909 43637 -13 20 37 70 3 53 -33 -33 3 53 0.00959 0.07025 -33.170 -----------------------------------------------------------------------------------------------------------------------------------------
Correction
Corrections
Q0: #24 80963 158230 -241 24 80963 158325 -241 Q3: #30 69446 43560 72 30 69446 43362 72
Quality check in X-Y plane after correction
Quality check in X-Y plane after correction Expand source
X-Y quality check for optical metrology measurements ---------------------------------------------------------------------------------------------------------------------------- segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) ---------------------------------------------------------------------------------------------------------------------------- segm: 0 20902 20906 -187 -183 43537 43539 85 87 -0.24346 48294 48299 -5 -4 -2 segm: 1 20904 20905 -136 -135 43541 43541 61 61 -0.17830 48295 48303 -7 -1 0 segm: 2 20905 20903 125 123 43539 43536 -60 -63 -0.16318 48294 48297 -3 2 3 segm: 3 20913 20903 126 116 43541 43536 -54 -59 -0.15923 48300 48297 2 10 5 segm: 4 20904 20902 3 1 43544 43544 0 0 0.00263 48302 48300 1 2 0 segm: 5 20908 20908 -291 -291 43542 43537 144 139 -0.38294 48302 48298 3 0 5 segm: 6 20906 20899 -24 -31 43538 43536 16 14 0.03619 48296 48293 3 7 2 segm: 7 20903 20900 111 108 43543 43536 -53 -60 -0.14410 48293 48300 -7 3 7 segm: 8 20910 20909 -89 -90 43543 43541 45 43 -0.11777 48303 48301 1 1 2 segm: 9 20908 20907 -19 -20 43550 43552 3 5 -0.02565 48304 48314 -9 1 -2 segm:10 20903 20907 0 4 43537 43536 2 1 -0.00263 48298 48293 4 -4 1 segm:11 20914 20912 76 74 43548 43545 -35 -38 -0.09868 48307 48308 0 2 3 segm:12 20905 20911 -144 -138 43551 43549 60 58 -0.18550 48301 48316 -15 -6 2 segm:13 20899 20894 -57 -62 43545 43544 35 34 -0.07829 48304 48293 10 5 1 segm:14 20876 20901 93 118 43540 43536 -61 -65 -0.13884 48278 48300 -22 -25 4 segm:15 20909 20903 -48 -54 43539 43538 27 26 0.06711 48299 48295 3 6 1 segm:16 20905 20903 87 85 43546 43545 -40 -41 0.11316 48303 48302 1 2 1 segm:17 20903 20906 -31 -28 43481 43486 6 11 -0.03887 48242 48252 -10 -3 -5 segm:18 20909 20910 90 91 43538 43541 -49 -46 -0.11909 48296 48303 -7 -1 -3 segm:19 20911 20924 -118 -105 43542 43547 46 51 0.14671 48303 48313 -9 -13 -5 segm:20 20907 20910 29 32 43550 43544 -9 -15 0.04013 48308 48304 4 -3 6 segm:21 20906 20909 -33 -30 43544 43545 13 14 -0.04145 48302 48305 -2 -3 -1 segm:22 20908 20908 2 2 43538 43541 -4 -1 -0.00263 48298 48300 -2 0 -3 segm:23 20912 20910 -110 -112 43542 43541 54 53 0.14606 48302 48302 0 2 1 segm:24 20912 20909 -36 -39 43545 43551 19 25 -0.04934 48311 48304 7 3 -6 segm:25 20905 20906 -108 -107 43545 43545 52 52 -0.14145 48303 48303 0 -1 0 segm:26 20890 20901 -215 -204 43531 43530 99 98 0.27575 48284 48288 -3 -11 1 segm:27 20910 20905 -253 -258 43539 43536 124 121 0.33624 48298 48298 0 5 3 segm:28 20905 20904 94 93 43540 43536 -43 -47 0.12305 48296 48296 0 1 4 segm:29 20897 20910 3 16 43542 43542 -1 -1 0.01250 48302 48296 6 -13 0 segm:30 20916 20908 -285 -293 43540 43537 143 140 0.38031 48303 48298 4 8 3 segm:31 20907 20911 -92 -88 43538 43538 46 46 0.11844 48301 48296 5 -4 0 ----------------------------------------------------------------------------------------------------------------------------
Constants
metrology processing command
> optical_metrology_cspad_v2_xpp -f 2023-02-22-CSPAD-CXI-SC3-Metrology.txt
Archive: /reg/g/psdm/detector/alignment/cspad/calib-cxi-SC3-2023-02-22/2023-02-22-CSPAD-CXI-SC3-geometry-v2.txt
References
- No labels
Overview
Content Tools