Back to CSPAD Alignment

Content

Optical measurements

Comments from Gabriel Blaj

From Blaj, Gabriel
Mon 6/26/2017 1:55 PM
To: Blaj, Gabriel;  Brown, Shaughnessy; Johnson, Tyler;  Hart, Philip Adam; Dubrovin, Mikhail;     Browne, Michael C.;

Hi,Update on MEC CSPAD Quad metrology
I sent Mikhail the latest metrology files for quads Q1 and Q3 (valid after 04/22/2017).
Q2 is being repaired (no new metrology).
Q0 still has the latest released metrology.

Before
 repairs, I also re-measured metrology data for Q1 and Q2 (valid before 
04/19/2017); they should match the latest released metrology.
Thanks,
Gabriel

 

Metrology files

They were combined in two files for entire CSPAD

2017-04-19-CSPAD-MEC-Before.txt, 2017-04-19-CSPAD-MEC-Before-corr.txt

  • Q0 - from previous ../calib-mec-2016-04-21/2016-04-21-CSPAD-MEC-Metrology-corr.txt
  • Q1 - MEC-Q1-Before-20170419.txt
  • Q2 - MEC-Q2-Before-20170419.txt
  • Q3 - MEC-Q3-After-20170422.txt  before is N/A

2017-04-22-CSPAD-MEC-After.txt

  • Q0 - from previous ../calib-mec-2016-04-21/2016-04-21-CSPAD-MEC-Metrology-corr.txt
  • Q1 - MEC-Q1-After-20170425.txt
  • Q2 - MEC-Q2-Before-20170419.txt after is N/A
  • Q3 - MEC-Q3-After-20170422.txt

SUMMARY OF DIFFERENCE between two files: Until Q2 is re-measured the difference between files is in Q1 ONLY!

 

Quality check in X-Y plane

File MEC-Q1-Before-20170419.txt - has 2 problematic segments:

2017-07-03 10:22:33 (info) CalibManager.GUIMetrology: Quality check in X-Y plane:
segm:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    angle(deg)   D1      D2      dD   d(dS)   d(dL)
Quad  1
segm: 4   20911   20514      85    -312    -43542  -43561      41      22    0.14932   48148   48303    -155     397      19
segm: 5   20904   20902      38      36    -43272  -43554     289       7   -0.04883   48066   48300    -234       2     282

with deviations of 397 and 282 um for two points, respectively.

Applied correction for two of measured points:

CORRECTION:
18c18
<     17  47134  66546      8
>     17  47134  66943      8
24c24
<     23  90435  69337    -74
>     23  90717  69337    -74

Corrected files

 

/reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/2017-04-19-CSPAD-MEC-Before-corr.txt
2017-07-03 11:01:07 (info) CalibManager.GUIMetrology: Quality check in X-Y plane:
segm:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    angle(deg)   D1      D2      dD   d(dS)   d(dL)
Quad  0
segm: 0  -20908  -20905     -87     -84     43544   43555     -46     -35   -0.11249   48308   48307       1      -3     -11
segm: 1  -20905  -20905     -43     -43     43550   43549     -17     -18   -0.05657   48310   48304       6       0       1
segm: 2  -20906  -20908     -18     -20    -43536  -43568      10     -22   -0.02500   48323   48296      27       2      32
segm: 3  -20907  -20918      39      28    -43540  -43546     -16     -22    0.04408   48307   48301       6      11       6
segm: 4   20917   20903     140     126    -43572  -43577      41      36   -0.17488   48354   48309      45      14       5
segm: 5   20905   20910    -172    -167    -43547  -43549     -79     -81    0.22301   48305   48308      -3      -5       2
segm: 6  -20927  -20903     -10      14    -43540  -43569      17     -12    0.00263   48312   48319      -7     -24      29
segm: 7  -20914  -20915      80      79    -43544  -43543     -39     -38    0.10461   48306   48305       1       1      -1
Quad  1
segm: 0  -20895  -20914     -63     -82     43527   43541     -48     -34   -0.09542   48287   48298     -11      19     -14
segm: 1  -20894  -20892     -11      -9     43534   43542      -9      -1   -0.01316   48291   48291       0      -2      -8
segm: 2  -20898  -20897    -152    -151    -43538  -43537      70      71   -0.19937   48295   48291       4      -1      -1
segm: 3  -20900  -20897      46      49    -43537  -43532     -28     -23    0.06251   48293   48288       5      -3      -5
segm: 4   20911   20911      85      85    -43542  -43561      41      22   -0.11182   48320   48303      17       0      19
segm: 5   20904   20902      38      36    -43554  -43554       7       7   -0.04867   48320   48300      20       2       0
segm: 6  -20923  -20939      35      19    -43583  -43574      -3       6    0.03550   48331   48357     -26      16      -9
segm: 7  -20892  -20888      72      76    -43530  -43505     -51     -26    0.09743   48274   48269       5      -4     -25
Quad  2
segm: 0  -20901  -20892      65      74     43531   43534      23      26    0.09147   48280   48296     -16      -9      -3
segm: 1  -20899  -20900      -1      -2     43542   43544      -7      -5   -0.00197   48294   48303      -9       1      -2
segm: 2  -20894  -20902     294     286    -43537  -43555    -133    -151    0.38156   48304   48299       5       8      18
segm: 3  -20901  -20894     148     155    -43535  -43536     -71     -72    0.19938   48290   48292      -2      -7       1
segm: 4   20897   20900    -251    -248    -43541  -43538    -122    -119    0.32833   48296   48295       1      -3      -3
segm: 5   20907   20903    -432    -436    -43545  -43549    -208    -212    0.57100   48308   48305       3       4       4
segm: 6  -20890  -20906     209     193    -43533  -43547     -84     -98    0.26450   48291   48300      -9      16      14
segm: 7  -20897  -20903     198     192    -43532  -43524    -104     -96    0.25668   48291   48280      11       6      -8
Quad  3
segm: 0  -20905  -20907    -228    -230     43542   43543    -113    -112   -0.30133   48299   48304      -5       2      -1
segm: 1  -20909  -20906    -106    -103     43545   43543     -50     -52   -0.13750   48302   48304      -2      -3       2
segm: 2  -20902  -20911    -180    -189    -43540  -43547      94      87   -0.24277   48301   48304      -3       9       7
segm: 3  -20913  -20912    -139    -138    -43542  -43550      71      63   -0.18223   48307   48307       0      -1       8
segm: 4   20915   20911      14      10    -43544  -43547      10       7   -0.01579   48304   48309      -5       4       3
segm: 5   20907   20917      21      31    -43542  -43574      36       4   -0.03420   48311   48324     -13     -10      32
segm: 6  -20900  -20912     -43     -55    -43543  -43546      26      23   -0.06447   48302   48303      -1      12       3
segm: 7  -20908  -20910      11       9    -43542  -43542      -4      -4    0.01316   48301   48302      -1       2       0

 

/reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/2017-04-22-CSPAD-MEC-After.txt

2017-07-03 10:16:03 (info) CalibManager.GUIMetrology: Quality check in X-Y plane:

segm:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    angle(deg)   D1      D2      dD   d(dS)   d(dL)
Quad  0
segm: 0  -20908  -20905     -87     -84     43544   43555     -46     -35   -0.11249   48308   48307       1      -3     -11
segm: 1  -20905  -20905     -43     -43     43550   43549     -17     -18   -0.05657   48310   48304       6       0       1
segm: 2  -20906  -20908     -18     -20    -43536  -43568      10     -22   -0.02500   48323   48296      27       2      32
segm: 3  -20907  -20918      39      28    -43540  -43546     -16     -22    0.04408   48307   48301       6      11       6
segm: 4   20917   20903     140     126    -43572  -43577      41      36   -0.17488   48354   48309      45      14       5
segm: 5   20905   20910    -172    -167    -43547  -43549     -79     -81    0.22301   48305   48308      -3      -5       2
segm: 6  -20927  -20903     -10      14    -43540  -43569      17     -12    0.00263   48312   48319      -7     -24      29
segm: 7  -20914  -20915      80      79    -43544  -43543     -39     -38    0.10461   48306   48305       1       1      -1
Quad  1
segm: 0  -20909  -20909     122     122     43542   43550      55      63    0.16052   48306   48305       1       0      -8
segm: 1  -20908  -20909     -22     -23     43544   43553     -12      -3   -0.02960   48310   48304       6       1      -9
segm: 2  -20906  -20916     -90    -100    -43541  -43545      53      49   -0.12500   48299   48308      -9      10       4
segm: 3  -20908  -20911      99      96    -43544  -43541     -53     -50    0.12830   48307   48298       9       3      -3
segm: 4   20909   20911    -103    -101    -43546  -43547     -48     -49    0.13421   48306   48307      -1      -2       1
segm: 5   20898   20909    -153    -142    -43545  -43547     -73     -75    0.19407   48306   48300       6     -11       2
segm: 6  -20906  -20909       7       4    -43569  -43547       6      28    0.00723   48298   48333     -35       3     -22
segm: 7  -20923  -20909      54      68    -43541  -43547     -22     -28    0.08026   48303   48310      -7     -14       6
Quad  2
segm: 0  -20901  -20892      65      74     43531   43534      23      26    0.09147   48280   48296     -16      -9      -3
segm: 1  -20899  -20900      -1      -2     43542   43544      -7      -5   -0.00197   48294   48303      -9       1      -2
segm: 2  -20894  -20902     294     286    -43537  -43555    -133    -151    0.38156   48304   48299       5       8      18
segm: 3  -20901  -20894     148     155    -43535  -43536     -71     -72    0.19938   48290   48292      -2      -7       1
segm: 4   20897   20900    -251    -248    -43541  -43538    -122    -119    0.32833   48296   48295       1      -3      -3
segm: 5   20907   20903    -432    -436    -43545  -43549    -208    -212    0.57100   48308   48305       3       4       4
segm: 6  -20890  -20906     209     193    -43533  -43547     -84     -98    0.26450   48291   48300      -9      16      14
segm: 7  -20897  -20903     198     192    -43532  -43524    -104     -96    0.25668   48291   48280      11       6      -8
Quad  3
segm: 0  -20905  -20907    -228    -230     43542   43543    -113    -112   -0.30133   48299   48304      -5       2      -1
segm: 1  -20909  -20906    -106    -103     43545   43543     -50     -52   -0.13750   48302   48304      -2      -3       2
segm: 2  -20902  -20911    -180    -189    -43540  -43547      94      87   -0.24277   48301   48304      -3       9       7
segm: 3  -20913  -20912    -139    -138    -43542  -43550      71      63   -0.18223   48307   48307       0      -1       8
segm: 4   20915   20911      14      10    -43544  -43547      10       7   -0.01579   48304   48309      -5       4       3
segm: 5   20907   20917      21      31    -43542  -43574      36       4   -0.03420   48311   48324     -13     -10      32
segm: 6  -20900  -20912     -43     -55    -43543  -43546      26      23   -0.06447   48302   48303      -1      12       3
segm: 7  -20908  -20910      11       9    -43542  -43542      -4      -4    0.01316   48301   48302      -1       2       0

Quality check in Z

/reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/2017-04-19-CSPAD-MEC-Before-corr.txt
 2017-07-03 11:01:07 (info) CalibManager.GUIMetrology: Quality check in Z:
segm:        SA      LA   XSize   YSize    dZS1  dZS2  dZL1  dZL2    dZSA  dZLA  ddZS  ddZL     dZX   dZY   angXZ(deg) angYZ(deg) dz3(um)
Quad  0
segm: 0  -20906   43549   43549   20906      24    25    36    37      24    36    -1    -1      36    24   0.04736    0.06578      0.994
segm: 1  -20905   43549   43549   20905      71    72   -10    -9      71    -9    -1    -1      -9    71  -0.01184    0.19459      1.000
segm: 2  -20907  -43552   20907   43552      29    29    19    19      29    19     0     0      29    19   0.07947    0.02500      0.017
segm: 3  -20912  -43543   20912   43543      27    -5    67    35      11    51    32    32      11    51   0.03014    0.06711     32.002
segm: 4   20910  -43574   43574   20910       2   -18     2   -18      -8    -8    20    20      -8    -8  -0.01052   -0.02192    -20.012
segm: 5   20907  -43548   43548   20907      60    -3    54    -9      28    22    63    63      22    28   0.02895    0.07673    -63.002
segm: 6  -20915  -43554   20915   43554       5     9    30    34       7    32    -4    -4       7    32   0.01918    0.04210     -3.988
segm: 7  -20914  -43543   20914   43543     -65   -63    24    26     -64    25    -2    -2     -64    25  -0.17533    0.03290     -2.004
Quad  1
segm: 0  -20904   43534   43534   20904       0     4   -77   -73       2   -75    -4    -4     -75     2  -0.09871    0.00548      4.024
segm: 1  -20893   43538   43538   20893      -8   -21   -57   -70     -14   -63    13    13     -63   -14  -0.08291   -0.03839    -12.988
segm: 2  -20897  -43537   20897   43537       9    51   -15    27      30     6   -42   -42      30     6   0.08225    0.00790    -42.003
segm: 3  -20898  -43534   20898   43534      27     9    -2   -20      18   -11    18    18      18   -11   0.04935   -0.01448     18.001
segm: 4   20911  -43551   43551   20911      21    13   102    94      17    98     8     8      98    17   0.12893    0.04658     -8.044
segm: 5   20903  -43554   43554   20903      19   -49    84    16     -15    50    68    68      50   -15   0.06578   -0.04112    -68.004
segm: 6  -20931  -43578   20931   43578     -78     6  -101   -17     -36   -59   -84   -84     -36   -59  -0.09855   -0.07757    -83.992
segm: 7  -20890  -43517   20890   43517      38    16    -3   -25      27   -14    22    22      27   -14   0.07405   -0.01843     22.011
Quad  2
segm: 0  -20896   43532   43532   20896     -35   -13     2    24     -24    13   -22   -22      13   -24   0.01711   -0.06581     21.983
segm: 1  -20899   43543   43543   20899     -45   -38    22    29     -41    25    -7    -7      25   -41   0.03290   -0.11240      7.001
segm: 2  -20898  -43546   20898   43546       9    16   -35   -28      12   -31    -7    -7      12   -31   0.03290   -0.04079     -7.005
segm: 3  -20897  -43535   20897   43535     -29   -11   -42   -24     -20   -33   -18   -18     -20   -33  -0.05484   -0.04343    -17.997
segm: 4   20898  -43539   43539   20898      21    11   -25   -35      16   -30    10    10     -30    16  -0.03948    0.04387    -10.003
segm: 5   20905  -43547   43547   20905     -46    -5   -31    10     -25   -10   -41   -41     -10   -25  -0.01316   -0.06852     41.002
segm: 6  -20898  -43540   20898   43540      10   -26    15   -21      -8    -3    36    36      -8    -3  -0.02193   -0.00395     35.973
segm: 7  -20900  -43528   20900   43528     -32   -48     2   -14     -40    -6    16    16     -40    -6  -0.10966   -0.00790     15.989
Quad  3
segm: 0  -20906   43542   43542   20906       0     5    -6    -1       2    -3    -5    -5      -3     2  -0.00395    0.00548      5.000
segm: 1  -20907   43544   43544   20907       1    25   -33    -9      13   -21   -24   -24     -21    13  -0.02763    0.03563     24.000
segm: 2  -20906  -43543   20906   43543       6     1    36    31       3    33     5     5       3    33   0.00822    0.04342      5.005
segm: 3  -20912  -43546   20912   43546       6   -21    39    12      -7    25    27    27      -7    25  -0.01918    0.03289     27.003
segm: 4   20913  -43545   43545   20913      20   -38    41   -17      -9    12    58    58      12    -9   0.01579   -0.02466    -58.010
segm: 5   20912  -43558   43558   20912     -11   -22    -1   -12     -16    -6    11    11      -6   -16  -0.00789   -0.04384    -10.989
segm: 6  -20906  -43544   20906   43544      14    15   -22   -21      14   -21    -1    -1      14   -21   0.03837   -0.02763     -0.993
segm: 7  -20909  -43542   20909   43542     -11    -3   -38   -30      -7   -34    -8    -8      -7   -34  -0.01918   -0.04474     -8.000

 

/reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/2017-04-22-CSPAD-MEC-After.txt

2017-07-03 10:16:03 (info) CalibManager.GUIMetrology: Quality check in Z:

segm:        SA      LA   XSize   YSize    dZS1  dZS2  dZL1  dZL2    dZSA  dZLA  ddZS  ddZL     dZX   dZY   angXZ(deg) angYZ(deg) dz3(um)
Quad  0
segm: 0  -20906   43549   43549   20906      24    25    36    37      24    36    -1    -1      36    24   0.04736    0.06578      0.994
segm: 1  -20905   43549   43549   20905      71    72   -10    -9      71    -9    -1    -1      -9    71  -0.01184    0.19459      1.000
segm: 2  -20907  -43552   20907   43552      29    29    19    19      29    19     0     0      29    19   0.07947    0.02500      0.017
segm: 3  -20912  -43543   20912   43543      27    -5    67    35      11    51    32    32      11    51   0.03014    0.06711     32.002
segm: 4   20910  -43574   43574   20910       2   -18     2   -18      -8    -8    20    20      -8    -8  -0.01052   -0.02192    -20.012
segm: 5   20907  -43548   43548   20907      60    -3    54    -9      28    22    63    63      22    28   0.02895    0.07673    -63.002
segm: 6  -20915  -43554   20915   43554       5     9    30    34       7    32    -4    -4       7    32   0.01918    0.04210     -3.988
segm: 7  -20914  -43543   20914   43543     -65   -63    24    26     -64    25    -2    -2     -64    25  -0.17533    0.03290     -2.004
Quad  1
segm: 0  -20909   43546   43546   20909     -22    -7   -43   -28     -14   -35   -15   -15     -35   -14  -0.04605   -0.03836     15.005
segm: 1  -20908   43548   43548   20908       6     2   -22   -26       4   -24     4     4     -24     4  -0.03158    0.01096     -3.995
segm: 2  -20911  -43543   20911   43543      15    35    -1    19      25     9   -20   -20      25     9   0.06850    0.01184    -19.981
segm: 3  -20909  -43542   20909   43542     -28    -6    23    45     -17    34   -22   -22     -17    34  -0.04658    0.04474    -22.004
segm: 4   20910  -43546   43546   20910      33   -40    67    -6      -3    30    73    73      30    -3   0.03947   -0.00822    -72.997
segm: 5   20903  -43546   43546   20903     -29   -13    32    48     -21    40   -16   -16      40   -21   0.05263   -0.05756     16.005
segm: 6  -20907  -43558   20907   43558      35    34   -22   -23      34   -22     1     1      34   -22   0.09318   -0.02894      1.016
segm: 7  -20916  -43544   20916   43544      -9    -1   -18   -10      -5   -14    -8    -8      -5   -14  -0.01370   -0.01842     -8.001
Quad  2
segm: 0  -20896   43532   43532   20896     -35   -13     2    24     -24    13   -22   -22      13   -24   0.01711   -0.06581     21.983
segm: 1  -20899   43543   43543   20899     -45   -38    22    29     -41    25    -7    -7      25   -41   0.03290   -0.11240      7.001
segm: 2  -20898  -43546   20898   43546       9    16   -35   -28      12   -31    -7    -7      12   -31   0.03290   -0.04079     -7.005
segm: 3  -20897  -43535   20897   43535     -29   -11   -42   -24     -20   -33   -18   -18     -20   -33  -0.05484   -0.04343    -17.997
segm: 4   20898  -43539   43539   20898      21    11   -25   -35      16   -30    10    10     -30    16  -0.03948    0.04387    -10.003
segm: 5   20905  -43547   43547   20905     -46    -5   -31    10     -25   -10   -41   -41     -10   -25  -0.01316   -0.06852     41.002
segm: 6  -20898  -43540   20898   43540      10   -26    15   -21      -8    -3    36    36      -8    -3  -0.02193   -0.00395     35.973
segm: 7  -20900  -43528   20900   43528     -32   -48     2   -14     -40    -6    16    16     -40    -6  -0.10966   -0.00790     15.989
Quad  3
segm: 0  -20906   43542   43542   20906       0     5    -6    -1       2    -3    -5    -5      -3     2  -0.00395    0.00548      5.000
segm: 1  -20907   43544   43544   20907       1    25   -33    -9      13   -21   -24   -24     -21    13  -0.02763    0.03563     24.000
segm: 2  -20906  -43543   20906   43543       6     1    36    31       3    33     5     5       3    33   0.00822    0.04342      5.005
segm: 3  -20912  -43546   20912   43546       6   -21    39    12      -7    25    27    27      -7    25  -0.01918    0.03289     27.003
segm: 4   20913  -43545   43545   20913      20   -38    41   -17      -9    12    58    58      12    -9   0.01579   -0.02466    -58.010
segm: 5   20912  -43558   43558   20912     -11   -22    -1   -12     -16    -6    11    11      -6   -16  -0.00789   -0.04384    -10.989
segm: 6  -20906  -43544   20906   43544      14    15   -22   -21      14   -21    -1    -1      14   -21   0.03837   -0.02763     -0.993
segm: 7  -20909  -43542   20909   43542     -11    -3   -38   -30      -7   -34    -8    -8      -7   -34  -0.01918   -0.04474     -8.000

 

Constants

(tick) Archive: /reg/g/psdm/detector/alignment/cspad/

  • /reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/calib/CsPad::CalibV1/MecTargetChamber.0:Cspad.0/geometry/0-end.data-2017-04-19-CSPAD-MEC-Before-corr
  • /reg/g/psdm/detector/alignment/cspad/calib-mec-2017-04-22/calib/CsPad::CalibV1/MecTargetChamber.0:Cspad.0/geometry/1-end.data-2017-04-22-CSPAD-MEC-After

References

 

 

 

  • No labels