Comments from Gabriel Blaj:
Attached, the latest CXI Camera2 metrology, to be used in all experiments performed AFTER 2016.05.25.
This replaces the previous version, 2016.02.05. Quads 0, 1 and 3 have been completely rebuilt; Quad 2 is unchanged.
As always, please use only Mikhail’s curated version; Excel files are known to contain bugs.
Thanks,
Gabriel
Metrology files:
- 2016-05-25-CSPAD-CXI-Camera2-Metrology.xlsx
- 2016-05-25-CSPAD-CXI-Camera2-Metrology.txt
- 2016-05-25-CSPAD-CXI-Camera2-Metrology-corr.txt
Quality check in X-Y plane
1st attempt:
segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 1 segm: 7 -20937 -21389 314 -138 -43569 -43543 53 79 0.11576 48327 48523 -196 452 -26
apply correction ~0.45mm for Quad 1 point 28:
< 28 65779 -384 121
> 28 65327 -384 121
segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 0 segm: 0 -20997 -20896 110 211 43550 43579 45 74 0.21109 48322 48354 -32 -101 -29 segm: 1 -20966 -20995 3 -26 43546 43596 -41 9 -0.01512 48349 48368 -19 29 -50 segm: 2 -20947 -20915 -5 27 -43550 -43552 3 1 0.01447 48313 48326 -13 -32 2 segm: 3 -20960 -20951 -57 -48 -43540 -43581 58 17 -0.06905 48327 48350 -23 -9 41 segm: 4 20921 20919 -3 -5 -43552 -43579 -4 -31 0.00526 48343 48316 27 2 27 segm: 5 20918 20916 1 -1 -43560 -43593 41 8 -0.00000 48314 48358 -44 2 33 segm: 6 -20951 -20925 -159 -133 -43552 -43590 120 82 -0.19199 48313 48368 -55 -26 38 segm: 7 -20919 -20911 119 127 -43549 -43553 -59 -63 0.16182 48314 48311 3 -8 4 Quad 1 segm: 0 -20943 -20949 -6 -12 43546 43543 -2 -5 -0.01184 48321 48319 2 6 3 segm: 1 -20908 -20917 -70 -79 43544 43545 -37 -36 -0.09803 48305 48306 -1 9 -1 segm: 2 -20938 -20924 -37 -23 -43577 -43542 -8 27 -0.03946 48331 48323 8 -14 -35 segm: 3 -20948 -20950 3 1 -43590 -43545 -47 -2 0.00263 48363 48321 42 2 -45 segm: 4 20966 20956 -39 -49 -43555 -43547 -33 -25 0.05789 48341 48325 16 10 -8 segm: 5 20926 20914 -29 -41 -43540 -43563 1 -22 0.04605 48309 48321 -12 12 23 segm: 6 -20927 -20918 17 26 -43556 -43549 -18 -11 0.02828 48321 48313 8 -9 -7 segm: 7 -20937 -20937 -138 -138 -43569 -43543 53 79 -0.18153 48327 48326 1 0 -26 Quad 2 segm: 0 -20944 -20953 91 82 43586 43585 40 39 0.11371 48356 48360 -4 9 1 segm: 1 -20973 -20937 -121 -85 43574 43547 -53 -80 -0.13548 48323 48354 -31 -36 27 segm: 2 -20911 -20953 10 -32 -43566 -43572 -7 -13 -0.01447 48350 48322 28 42 6 segm: 3 -20908 -20909 -52 -53 -43544 -43546 27 25 -0.06908 48303 48305 -2 1 2 segm: 4 20947 20911 58 22 -43547 -43547 9 9 -0.05263 48324 48306 18 36 0 segm: 5 20917 20911 -86 -92 -43548 -43554 -38 -44 0.11709 48310 48314 -4 6 6 segm: 6 -20908 -20907 66 67 -43545 -43572 -26 -53 0.08747 48323 48309 14 -1 27 segm: 7 -20913 -20909 -46 -42 -43483 -43480 23 26 -0.05798 48245 48251 -6 -4 -3 Quad 3 segm: 0 -20999 -20955 35 79 43543 43569 38 64 0.07498 48365 48323 42 -44 -26 segm: 1 -20907 -20905 -52 -50 43564 43549 -10 -25 -0.06709 48320 48307 13 -2 15 segm: 2 -20936 -20910 -92 -66 -43544 -43576 48 16 -0.10391 48329 48319 10 -26 32 segm: 3 -20932 -20904 -210 -182 -43551 -43555 92 88 -0.25784 48320 48312 8 -28 4 segm: 4 20908 20949 1 42 -43545 -43550 -9 -14 -0.02829 48335 48297 38 -41 5 segm: 5 20936 20931 -85 -90 -43560 -43546 -49 -35 0.11511 48322 48322 0 5 -14 segm: 6 -20944 -20945 38 37 -43578 -43521 -70 -13 0.04934 48345 48303 42 1 -57 segm: 7 -20919 -20942 90 67 -43563 -43553 -41 -31 0.10326 48324 48327 -3 23 -10
Quality check in Z
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) Quad 0 segm: 0 -20946 43564 43564 20946 20 20 -8 -8 20 -8 0 0 -8 20 -0.01052 0.05471 0.101 segm: 1 -20980 43571 43571 20980 13 22 42 51 17 46 -9 -9 46 17 0.06049 0.04643 8.923 segm: 2 -20931 -43551 20931 43551 -51 -36 65 80 -43 72 -15 -15 -43 72 -0.11771 0.09472 -14.941 segm: 3 -20955 -43560 20955 43560 -43 -42 43 44 -42 43 -1 -1 -42 43 -0.11484 0.05656 -0.940 segm: 4 20920 -43565 43565 20920 474 338 25 -111 406 -43 136 136 -43 406 -0.05655 1.11181 -135.965 segm: 5 20917 -43576 43576 20917 32 23 -74 -83 27 -78 9 9 -78 27 -0.10256 0.07396 -8.942 segm: 6 -20938 -43571 20938 43571 -59 -37 -67 -45 -48 -56 -22 -22 -48 -56 -0.13135 -0.07364 -21.993 segm: 7 -20915 -43551 20915 43551 -30 -14 -11 5 -22 -3 -16 -16 -22 -3 -0.06027 -0.00395 -15.994 Quad 1 segm: 0 -20946 43544 43544 20946 -6 -5 140 141 -5 140 -1 -1 140 -5 0.18421 -0.01368 1.011 segm: 1 -20912 43544 43544 20912 13 -10 147 124 1 135 23 23 135 1 0.17763 0.00274 -23.008 segm: 2 -20931 -43559 20931 43559 -39 -16 123 146 -27 134 -23 -23 -27 134 -0.07391 0.17626 -23.107 segm: 3 -20949 -43567 20949 43567 -34 -16 167 185 -25 176 -18 -18 -25 176 -0.06838 0.23146 -18.193 segm: 4 20961 -43551 43551 20961 386 88 263 -35 237 114 298 298 114 237 0.14998 0.64780 -297.902 segm: 5 20920 -43551 43551 20920 -35 15 -38 12 -10 -13 -50 -50 -13 -10 -0.01710 -0.02739 50.029 segm: 6 -20922 -43552 20922 43552 12 -60 66 -6 -24 30 72 72 -24 30 -0.06572 0.03947 72.026 segm: 7 -20937 -43556 20937 43556 -71 -40 -19 12 -55 -3 -31 -31 -55 -3 -0.15051 -0.00395 -31.007 Quad 2 segm: 0 -20948 43585 43585 20948 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 1 -20955 43560 43560 20955 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 2 -20932 -43569 20932 43569 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 3 -20908 -43545 20908 43545 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 4 20929 -43547 43547 20929 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 5 20914 -43551 43551 20914 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 6 -20907 -43558 20907 43558 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 7 -20911 -43481 20911 43481 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 Quad 3 segm: 0 -20977 43556 43556 20977 6 6 -94 -94 6 -94 0 0 -94 6 -0.12365 0.01639 0.069 segm: 1 -20906 43556 43556 20906 -14 -10 -14 -10 -12 -12 -4 -4 -12 -12 -0.01579 -0.03289 3.995 segm: 2 -20923 -43560 20923 43560 14 8 3 -3 11 0 6 6 11 0 0.03012 0.00000 5.988 segm: 3 -20918 -43553 20918 43553 16 -17 -3 -36 0 -19 33 33 0 -19 0.00000 -0.02500 33.020 segm: 4 20928 -43547 43547 20928 92 227 -254 -119 159 -186 -135 -135 -186 159 -0.24472 0.43529 134.575 segm: 5 20933 -43553 43553 20933 44 42 -87 -89 43 -88 2 2 -88 43 -0.11577 0.11770 -2.018 segm: 6 -20944 -43549 20944 43549 23 15 -104 -112 19 -108 8 8 19 -108 0.05198 -0.14209 8.147 segm: 7 -20930 -43558 20930 43558 3 8 -29 -24 5 -26 -5 -5 5 -26 0.01369 -0.03420 -4.986
Constants
Archive: /reg/g/psdm/detector/alignment/cspad/calib-cxi-camera2-2016-05-25/calib/CsPad::CalibV1/CxiDs2.0:Cspad.0/geometry/0-end.data