Optical measurements
Comments from Gabriel Blaj:
These replace the May 2014 versions and should be used for all experiments after September 24, 2014.
Changes in CXI-DS2:
- Quad 1 (zero based counting) reworked; metrology updated in the attached file: DS2-Metrology-Aug-27-2014
- Q1 Z values: measured incorrectly (OGP bug?), please discard.
- Mikhail, could you please check Q1 values for consistency?
- The other 3 quads (Q0, Q2, Q3) are unchanged (as in the previous metrology: May 14, 2014).
- Please note that this camera has been renamed. Possible new name: "CXI Camera 2"+ version number. Jason, could you please confirm the exact name and version.
As always, these excel metrology files are NOT FOR DISTRIBUTION to avoid user confusion. Only Mikhail's curated version should be distributed and used.
Metrology files:
- 2014-08-27-CSPAD-CXI-DS2-Metrology.xlsx
- 2014-08-27-CSPAD-CXI-DS2-Metrology.txt
- 2014-08-27-CSPAD-CXI-DS2-Metrology-disc-Z.txt
- Z coordinates for Q0, Q2, Q3 were not measured
- Q1 Z values: measured incorrectly - discard
No correction for X-Y, Z-data for Q1 is discarded
Quality check in X-Y plane
Quality check in XY plane: segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 0 segm: 0 -20937 -20905 101 133 43556 43544 59 47 0.15393 48311 48317 -6 -32 12 segm: 1 -20911 -20905 65 71 43571 43551 31 11 0.08944 48308 48329 -21 -6 20 segm: 2 -20927 -20910 51 68 -43542 -43548 -17 -23 0.07829 48301 48316 -15 -17 6 segm: 3 -20931 -20916 31 46 -43587 -43591 3 -1 0.05061 48333 48368 -35 -15 4 segm: 4 20943 20946 55 58 -43546 -43544 27 29 -0.07434 48319 48320 -1 -3 -2 segm: 5 20916 20903 138 125 -43558 -43543 58 73 -0.17300 48308 48312 -4 13 -15 segm: 6 -20948 -20942 -58 -52 -43574 -43510 -34 30 -0.07237 48343 48292 51 -6 -64 segm: 7 -20915 -20941 2 -24 -43571 -43550 6 27 -0.01447 48316 48337 -21 26 -21 Quad 1 segm: 0 -20932 -20954 -138 -160 43566 43569 -61 -58 -0.19595 48351 48329 22 22 -3 segm: 1 -20948 -20950 46 44 43558 43548 26 16 0.05920 48328 48330 -2 2 10 segm: 2 -20930 -20927 64 67 -43572 -43542 -54 -24 0.08616 48332 48317 15 -3 -30 segm: 3 -20951 -20912 95 134 -43592 -43547 -97 -52 0.15057 48354 48319 35 -39 -45 segm: 4 20882 20956 -74 0 -43558 -43549 -46 -37 0.04867 48338 48296 42 -74 -9 segm: 5 20926 20925 -51 -52 -43552 -43577 -2 -27 0.06773 48320 48339 -19 1 25 segm: 6 -20924 -20921 91 94 -43563 -43566 -42 -45 0.12166 48327 48329 -2 -3 3 segm: 7 -20940 -20912 -95 -67 -43563 -43547 29 45 -0.10655 48323 48319 4 -28 -16 Quad 2 segm: 0 -20944 -20953 91 82 43586 43585 40 39 0.11371 48356 48360 -4 9 1 segm: 1 -20973 -20937 -121 -85 43574 43547 -53 -80 -0.13548 48323 48354 -31 -36 27 segm: 2 -20911 -20953 10 -32 -43566 -43572 -7 -13 -0.01447 48350 48322 28 42 6 segm: 3 -20908 -20909 -52 -53 -43544 -43546 27 25 -0.06908 48303 48305 -2 1 2 segm: 4 20947 20911 58 22 -43547 -43547 9 9 -0.05263 48324 48306 18 36 0 segm: 5 20917 20911 -86 -92 -43548 -43554 -38 -44 0.11709 48310 48314 -4 6 6 segm: 6 -20908 -20907 66 67 -43545 -43572 -26 -53 0.08747 48323 48309 14 -1 27 segm: 7 -20913 -20909 -46 -42 -43483 -43480 23 26 -0.05798 48245 48251 -6 -4 -3 Quad 3 segm: 0 -20998 -20955 85 128 43540 43537 61 58 0.14015 48335 48320 15 -43 3 segm: 1 -20906 -20904 -2 0 43566 43544 20 -2 -0.00132 48320 48303 17 -2 22 segm: 2 -20911 -20910 17 18 -43544 -43579 -3 -38 0.02302 48331 48309 22 -1 35 segm: 3 -20932 -20903 -98 -69 -43548 -43559 45 34 -0.10985 48316 48315 1 -29 11 segm: 4 20967 20948 -18 -37 -43544 -43548 -22 -26 0.03618 48336 48317 19 19 4 segm: 5 20935 20929 -110 -116 -43569 -43549 -60 -40 0.14864 48323 48331 -8 6 -20 segm: 6 -20911 -20913 32 30 -43547 -43547 -13 -13 0.04079 48306 48309 -3 2 0 segm: 7 -20924 -20909 91 106 -43561 -43554 -43 -36 0.12957 48312 48326 -14 -15 -7
Quality check in Z
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) Quad 0 segm: 0 -20921 43550 43550 20921 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 1 -20908 43561 43561 20908 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 2 -20918 -43545 20918 43545 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 3 -20923 -43589 20923 43589 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 4 20944 -43545 43545 20944 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 5 20909 -43550 43550 20909 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 6 -20945 -43542 20945 43542 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 7 -20928 -43560 20928 43560 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 Quad 1 segm: 0 -20943 43567 43567 20943 -97 6 173 276 -45 224 -103 -103 224 -45 0.29458 -0.12311 103.079 segm: 1 -20949 43553 43553 20949 -37 -47 110 100 -42 105 10 10 105 -42 0.13813 -0.11487 -9.973 segm: 2 -20928 -43557 20928 43557 371 -7 417 39 182 228 378 378 182 228 0.49826 0.29991 377.974 segm: 3 -20931 -43569 20931 43569 -38 21 25 84 -8 54 -59 -59 -8 54 -0.02190 0.07101 -59.126 segm: 4 20919 -43553 43553 20919 33 189 -242 -86 111 -164 -156 -156 -164 111 -0.21575 0.30402 155.273 segm: 5 20925 -43564 43564 20925 37 24 -138 -151 30 -144 13 13 -144 30 -0.18939 0.08214 -12.920 segm: 6 -20922 -43564 20922 43564 314 -44 252 -106 135 73 358 358 135 73 0.36970 0.09601 357.997 segm: 7 -20926 -43555 20926 43555 -61 -21 -131 -91 -41 -111 -40 -40 -41 -111 -0.11226 -0.14602 -39.938 Quad 2 segm: 0 -20948 43585 43585 20948 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 1 -20955 43560 43560 20955 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 2 -20932 -43569 20932 43569 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 3 -20908 -43545 20908 43545 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 4 20929 -43547 43547 20929 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 5 20914 -43551 43551 20914 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 6 -20907 -43558 20907 43558 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 7 -20911 -43481 20911 43481 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 Quad 3 segm: 0 -20976 43538 43538 20976 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 1 -20905 43555 43555 20905 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 2 -20910 -43561 20910 43561 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 3 -20917 -43553 20917 43553 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 4 20957 -43546 43546 20957 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 5 20932 -43559 43559 20932 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 6 -20912 -43547 20912 43547 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 7 -20916 -43557 20916 43557 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000
Constants
center/0-end.data (pixel)
197.75 198.00 301.45 95.97 623.79 624.47 707.80 499.34 198.29 197.99 303.05 95.42 623.88 623.49 703.30 498.56 197.99 198.38 302.65 96.30 623.07 622.57 703.18 497.79 197.11 198.00 302.55 96.64 623.23 623.09 709.28 496.75 299.40 95.76 621.88 621.77 517.33 724.11 196.36 196.69 303.80 95.49 623.17 621.60 517.71 722.53 194.55 194.38 302.90 94.91 621.91 621.75 518.61 722.37 197.52 197.22 299.32 95.08 624.14 623.75 516.82 722.27 195.66 195.69 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00
or the same in um
21736 21764 33135 10548 68567 68641 77801 54887 21796 21763 33311 10488 68577 68533 77306 54801 21762 21805 33267 10584 68488 68432 77293 54717 21666 21764 33256 10622 68505 68490 77964 54602 32910 10525 68357 68345 56864 79593 21584 21619 33393 10496 68498 68326 56906 79420 21384 21366 33294 10433 68360 68343 57005 79403 21711 21678 32901 10451 68605 68562 56808 79392 21507 21509 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
tilt/0-end.data (degree)
0.15393 0.08944 0.07829 0.05061 -0.07434 -0.17300 -0.07237 -0.01447 -0.19595 0.05920 0.08616 0.15057 0.04867 0.06773 0.12166 -0.10655 0.11371 -0.13548 -0.01447 -0.06908 -0.05263 0.11709 0.08747 -0.05798 0.14015 -0.00132 0.02302 -0.10985 0.03618 0.14864 0.04079 0.12957
geometry/0-end.data
# TITLE Geometry parameters of CSPAD-CXI # DATE_TIME 2014-10-28 12:55:54 PDT # METROLOGY /reg/neh/home1/dubrovin/LCLS/CSPadMetrologyProc/2014-08-27-CSPAD-CXI-DS2-Metrology.txt # AUTHOR dubrovin # EXPERIMENT Any # DETECTOR CSPAD-CXI # CALIB_TYPE geometry # COMMENT:01 Table contains the list of geometry parameters for alignment of 2x1 sensors, quads, CSPAD, etc # COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame # PARAM:01 PARENT - name and version of the parent object # PARAM:02 PARENT_IND - index of the parent object # PARAM:03 OBJECT - name and version of the object # PARAM:04 OBJECT_IND - index of the new object # PARAM:05 X0 - x-coordinate [um] of the object origin in the parent frame # PARAM:06 Y0 - y-coordinate [um] of the object origin in the parent frame # PARAM:07 Z0 - z-coordinate [um] of the object origin in the parent frame # PARAM:08 ROT_Z - object design rotation angle [deg] around Z axis of the parent frame # PARAM:09 ROT_Y - object design rotation angle [deg] around Y axis of the parent frame # PARAM:10 ROT_X - object design rotation angle [deg] around X axis of the parent frame # PARAM:11 TILT_Z - object tilt angle [deg] around Z axis of the parent frame # PARAM:12 TILT_Y - object tilt angle [deg] around Y axis of the parent frame # PARAM:13 TILT_X - object tilt angle [deg] around X axis of the parent frame # HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X QUAD:V1 0 SENS2X1:V1 0 21736 32910 0 0 0 0 0.15393 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 1 21764 10525 0 0 0 0 0.08944 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 2 33135 68357 0 270 0 0 0.07829 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 3 10548 68345 0 270 0 0 0.05061 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 4 68567 56864 0 180 0 0 -0.07434 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 5 68641 79593 0 180 0 0 -0.17300 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 6 77801 21584 0 270 0 0 -0.07237 0.00000 0.00000 QUAD:V1 0 SENS2X1:V1 7 54887 21619 0 270 0 0 -0.01447 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 0 21796 33393 0 0 0 0 -0.19595 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 1 21763 10496 0 0 0 0 0.05920 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 2 33311 68498 0 270 0 0 0.08616 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 3 10488 68326 0 270 0 0 0.15057 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 4 68577 56906 0 180 0 0 0.04867 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 5 68533 79420 0 180 0 0 0.06773 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 6 77306 21384 0 270 0 0 0.12166 0.00000 0.00000 QUAD:V1 1 SENS2X1:V1 7 54801 21366 0 270 0 0 -0.10655 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 0 21762 33294 0 0 0 0 0.11371 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 1 21805 10433 0 0 0 0 -0.13548 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 2 33267 68360 0 270 0 0 -0.01447 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 3 10584 68343 0 270 0 0 -0.06908 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 4 68488 57005 0 180 0 0 -0.05263 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 5 68432 79403 0 180 0 0 0.11709 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 6 77293 21711 0 270 0 0 0.08747 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 7 54717 21678 0 270 0 0 -0.05798 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 0 21666 32901 0 0 0 0 0.14015 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 1 21764 10451 0 0 0 0 -0.00132 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 2 33256 68605 0 270 0 0 0.02302 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 3 10622 68562 0 270 0 0 -0.10985 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 4 68505 56808 0 180 0 0 0.03618 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 5 68490 79392 0 180 0 0 0.14864 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 6 77964 21507 0 270 0 0 0.04079 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 7 54602 21509 0 270 0 0 0.12957 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 0 -4500 -4500 0 90 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 1 -4500 4500 0 0 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 2 4500 4500 0 270 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 3 4500 -4500 0 180 0 0 0.00000 0.00000 0.00000 RAIL 0 CSPAD:V1 0 0 0 1000000 0 0 0 0.00000 0.00000 0.00000 IP 0 RAIL 0 0 0 0 0 0 0 0.00000 0.00000 0.00000
Quads alignment - TBD
Alignment from previous (2014-05-15) run data exp=cxid9114:run=170
marg_gap_shift/0-end.data
15 40 4 40 15 40 4 40 0 0 0 0
offset/0-end.data
0 0 820 820 0 820 820 0 0 0 0 0
offset_corr/0-end.data
0 -12 -3 6 0 9 9 4 0 0 0 0
offset_corr/1-end.data
0 -8 -1 6 0 8 6 3 0 0 0 0