Content

 

Optical measurements

Comments from Gabriel Blaj:

DS2 (also called DSD) was completely rebuilt; all 4 quads are different. Please process / clean up all of them.

These metrology files should be used for all experiments AFTER May 15th (until next upgrade/repair).

Metrology files:

Z-information is not available!

 

Quality check in X-Y plane

 

 segm:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    angle(deg)   D1      D2      dD   d(dS)   d(dL)
Quad  0
segm: 0   20937   20905     133     101     43544   43556     -59     -47    0.15393   48317   48311       6      32     -12
segm: 1   20911   20905      71      65     43551   43571     -31     -11    0.08944   48329   48308      21       6     -20
segm: 2   20910   20927      68      51     43548   43542      23      17    0.07763   48301   48316     -15      17       6
segm: 3   20916   20931      46      31     43591   43587       1      -3    0.04995   48333   48368     -35      15       4
segm: 4   20946   20943     -55     -58     43546   43544      29      27   -0.07368   48320   48319       1       3       2
segm: 5   20903   20916    -138    -125     43558   43543      73      58   -0.17235   48312   48308       4     -13      15
segm: 6   20942   20948     -52     -58     43510   43574     -30      34   -0.07237   48343   48292      51       6     -64
segm: 7   20941   20915     -24       2     43550   43571     -27      -6   -0.01447   48316   48337     -21     -26     -21

Quad  1
segm: 0   20930   20960    -260    -230     43421   43416     105     100   -0.32331   48193   48221     -28     -30       5
segm: 1   20949   20955     -44     -38     43546   43562      17      33   -0.05394   48336   48326      10      -6     -16
segm: 2   20924   20936     -17     -29     43540   43572     -20      12   -0.03026   48330   48317      13      12     -32
segm: 3   20913   20952      46       7     43547   43592      10      55    0.03419   48354   48319      35      39     -45
segm: 4   20957   20942     -15     -30     43562   43553       5      -4   -0.02894   48324   48342     -18      15       9
segm: 5   20922   20924     -37     -35     43550   43569      19      38   -0.04735   48334   48313      21      -2     -19
segm: 6   20906   20964     306     248     43532   43556     120     144    0.36448   48315   48317      -2      58     -24
segm: 7   20939   20946     -41     -48     43546   43553     -16      -9   -0.05789   48331   48315      16       7      -7

Quad  2
segm: 0   20944   20953      82      91     43585   43586     -40     -39    0.11305   48360   48356       4      -9      -1
segm: 1   20973   20937     -85    -121     43547   43574      53      80   -0.13548   48354   48323      31      36     -27
segm: 2   20953   20911     -32      10     43572   43566      13       7   -0.01447   48350   48322      28     -42       6
segm: 3   20909   20908     -53     -52     43546   43544     -25     -27   -0.06842   48303   48305      -2      -1       2
segm: 4   20911   20947     -58     -22     43547   43547       9       9   -0.05263   48306   48324     -18     -36       0
segm: 5   20911   20917      86      92     43548   43554     -44     -38    0.11709   48314   48310       4      -6      -6
segm: 6   20907   20908      67      66     43572   43545      53      26    0.08682   48323   48309      14       1      27
segm: 7   20909   20913     -42     -46     43480   43483     -26     -23   -0.05798   48245   48251      -6       4      -3

Quad  3
segm: 0   20998   20955     128      85     43537   43540     -61     -58    0.13950   48320   48335     -15      43      -3
segm: 1   20906   20904       0      -2     43544   43566     -20       2   -0.00132   48303   48320     -17       2     -22
segm: 2   20910   20911      18      17     43579   43544      38       3    0.02236   48331   48309      22       1      35
segm: 3   20903   20932     -69     -98     43559   43548     -34     -45   -0.10919   48316   48315       1      29      11
segm: 4   20948   20967      18      37     43544   43548     -26     -22    0.03553   48317   48336     -19     -19      -4
segm: 5   20929   20935     110     116     43569   43549     -40     -60    0.14864   48331   48323       8      -6      20
segm: 6   20913   20911      30      32     43547   43547      13      13    0.04079   48306   48309      -3      -2       0
segm: 7   20909   20924     106      91     43554   43561      36      43    0.12891   48312   48326     -14      15      -7

Quality check in Z

 

 segm:        SA      LA   XSize   YSize    dZS1  dZS2  dZL1  dZL2    dZSA  dZLA  ddZS  ddZL     dZX   dZY   angXZ(deg) angYZ(deg) dz3(um)
Quad  0
segm: 0   20921   43550   43550   20921       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 1   20908   43561   43561   20908       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 2   20918   43545   20918   43545       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 3   20923   43589   20923   43589       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 4   20944   43545   43545   20944       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 5   20909   43550   43550   20909       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 6   20945   43542   20945   43542       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 7   20928   43560   20928   43560       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000

Quad  1
segm: 0   20945   43418   43418   20945       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 1   20952   43554   43554   20952       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 2   20930   43556   20930   43556       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 3   20932   43569   20932   43569       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 4   20949   43557   43557   20949       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 5   20923   43559   43559   20923       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 6   20935   43544   20935   43544       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 7   20942   43549   20942   43549       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000

Quad  2
segm: 0   20948   43585   43585   20948       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 1   20955   43560   43560   20955       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 2   20932   43569   20932   43569       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 3   20908   43545   20908   43545       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 4   20929   43547   43547   20929       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 5   20914   43551   43551   20914       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 6   20907   43558   20907   43558       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 7   20911   43481   20911   43481       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000

Quad  3
segm: 0   20976   43538   43538   20976       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 1   20905   43555   43555   20905       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 2   20910   43561   20910   43561       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 3   20917   43553   20917   43553       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 4   20957   43546   43546   20957       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 5   20932   43559   43559   20932       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 6   20912   43547   20912   43547       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000
segm: 7   20916   43557   20916   43557       0     0     0     0       0     0     0     0       0     0   0.00000    0.00000      0.000

 

 

(tick) center/0-end.data (pixel)

 197.75   198.00   301.45    95.97   623.79   624.47   707.80   499.34  
 198.24   198.19   304.34    96.68   624.95   624.94   705.13   498.42  
 197.99   198.38   302.65    96.30   623.07   622.57   703.18   497.79  
 197.11   198.00   302.55    96.64   623.23   623.09   709.28   496.75  

 299.40    95.76   621.88   621.77   517.33   724.11   196.36   196.69  
 303.37    95.08   622.53   621.39   516.56   721.28   191.88   191.55  
 302.90    94.91   621.91   621.75   518.61   722.37   197.52   197.22  
 299.32    95.08   624.14   623.75   516.82   722.27   195.66   195.69  

   0.00     0.00     0.00     0.00     0.00     0.00     0.00     0.00  
   0.00     0.00     0.00     0.00     0.00     0.00     0.00     0.00  
   0.00     0.00     0.00     0.00     0.00     0.00     0.00     0.00  
   0.00     0.00     0.00     0.00     0.00     0.00     0.00     0.00

 

or the same in um

 21736   21764   33135   10548   68567   68641   77801   54887  
 21790   21785   33453   10627   68694   68693   77507   54786  
 21762   21805   33267   10584   68488   68432   77293   54717  
 21666   21764   33256   10622   68505   68490   77964   54602  

 32910   10525   68357   68345   56864   79593   21584   21619  
 33346   10451   68428   68302   56780   79283   21091   21054  
 33294   10433   68360   68343   57005   79403   21711   21678  
 32901   10451   68605   68562   56808   79392   21507   21509  

     0       0       0       0       0       0       0       0  
     0       0       0       0       0       0       0       0  
     0       0       0       0       0       0       0       0  
     0       0       0       0       0       0       0       0 

 

(tick) tilt/0-end.data (degree)

 0.15393   0.08944   0.07763   0.04995  -0.07368  -0.17235  -0.07237  -0.01447  
-0.32331  -0.05394  -0.03026   0.03419  -0.02894  -0.04735   0.36448  -0.05789  
 0.11305  -0.13548  -0.01447  -0.06842  -0.05263   0.11709   0.08682  -0.05798  
 0.13950  -0.00132   0.02236  -0.10919   0.03553   0.14864   0.04079   0.12891

 

(tick) geometry/0-end.data

# TITLE      Geometry parameters of CSPAD-CXI
# DATE_TIME  2014-10-03 12:12:05 PDT
# METROLOGY  /reg/neh/home1/dubrovin/LCLS/CSPadMetrologyProc/2014-05-15-CSPAD-CXI-DS2-Metrology.txt
# AUTHOR     dubrovin
# EXPERIMENT Any
# DETECTOR   CSPAD-CXI
# CALIB_TYPE geometry
# COMMENT:01 Table contains the list of geometry parameters for alignment of 2x1 sensors, quads, CSPAD, etc
# COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame
# PARAM:01 PARENT     - name and version of the parent object
# PARAM:02 PARENT_IND - index of the parent object
# PARAM:03 OBJECT     - name and version of the object
# PARAM:04 OBJECT_IND - index of the new object
# PARAM:05 X0         - x-coordinate [um] of the object origin in the parent frame
# PARAM:06 Y0         - y-coordinate [um] of the object origin in the parent frame
# PARAM:07 Z0         - z-coordinate [um] of the object origin in the parent frame
# PARAM:08 ROT_Z      - object design rotation angle [deg] around Z axis of the parent frame
# PARAM:09 ROT_Y      - object design rotation angle [deg] around Y axis of the parent frame
# PARAM:10 ROT_X      - object design rotation angle [deg] around X axis of the parent frame
# PARAM:11 TILT_Z     - object tilt angle [deg] around Z axis of the parent frame
# PARAM:12 TILT_Y     - object tilt angle [deg] around Y axis of the parent frame
# PARAM:13 TILT_X     - object tilt angle [deg] around X axis of the parent frame

# HDR PARENT IND        OBJECT IND    X0[um]  Y0[um]  Z0[um]   ROT-Z ROT-Y ROT-X     TILT-Z   TILT-Y   TILT-X

QUAD:V1        0  SENS2X1:V1     0     21736   32910       0       0     0     0    0.15393  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     1     21764   10525       0       0     0     0    0.08944  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     2     33135   68357       0     270     0     0    0.07829  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     3     10548   68345       0     270     0     0    0.05061  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     4     68567   56864       0     180     0     0   -0.07434  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     5     68641   79593       0     180     0     0   -0.17300  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     6     77801   21584       0     270     0     0   -0.07237  0.00000  0.00000 
QUAD:V1        0  SENS2X1:V1     7     54887   21619       0     270     0     0   -0.01447  0.00000  0.00000 

QUAD:V1        1  SENS2X1:V1     0     21790   33346       0       0     0     0   -0.32330  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     1     21785   10451       0       0     0     0   -0.05394  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     2     33453   68428       0     270     0     0   -0.03026  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     3     10627   68302       0     270     0     0    0.03485  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     4     68694   56780       0     180     0     0   -0.02960  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     5     68693   79283       0     180     0     0   -0.04735  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     6     77507   21091       0     270     0     0    0.36448  0.00000  0.00000 
QUAD:V1        1  SENS2X1:V1     7     54786   21054       0     270     0     0   -0.05855  0.00000  0.00000 

QUAD:V1        2  SENS2X1:V1     0     21762   33294       0       0     0     0    0.11371  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     1     21805   10433       0       0     0     0   -0.13548  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     2     33267   68360       0     270     0     0   -0.01447  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     3     10584   68343       0     270     0     0   -0.06908  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     4     68488   57005       0     180     0     0   -0.05263  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     5     68432   79403       0     180     0     0    0.11709  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     6     77293   21711       0     270     0     0    0.08747  0.00000  0.00000 
QUAD:V1        2  SENS2X1:V1     7     54717   21678       0     270     0     0   -0.05798  0.00000  0.00000 

QUAD:V1        3  SENS2X1:V1     0     21666   32901       0       0     0     0    0.14015  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     1     21764   10451       0       0     0     0   -0.00132  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     2     33256   68605       0     270     0     0    0.02302  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     3     10622   68562       0     270     0     0   -0.10985  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     4     68505   56808       0     180     0     0    0.03618  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     5     68490   79392       0     180     0     0    0.14864  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     6     77964   21507       0     270     0     0    0.04079  0.00000  0.00000 
QUAD:V1        3  SENS2X1:V1     7     54602   21509       0     270     0     0    0.12957  0.00000  0.00000 

CSPAD:V1       0  QUAD:V1        0     -4500   -4500       0      90     0     0    0.00000  0.00000  0.00000 
CSPAD:V1       0  QUAD:V1        1     -4500    4500       0       0     0     0    0.00000  0.00000  0.00000 
CSPAD:V1       0  QUAD:V1        2      4500    4500       0     270     0     0    0.00000  0.00000  0.00000 
CSPAD:V1       0  QUAD:V1        3      4500   -4500       0     180     0     0    0.00000  0.00000  0.00000 

 

2014-07-09 Quads alignment

Use data exp=cxid9114:run=170

 

(tick) marg_gap_shift/0-end.data

15   40   4  40
15   40   4  40
 0    0   0   0

 

(tick) offset/0-end.data

  0    0  820  820
  0  820  820    0
  0    0    0    0 

 

(tick) offset_corr/0-end.data

  0  -12  -3   6
  0    9   9   4
  0    0   0   0

 

Plots for alignment

Circle radius [pix] = 418

For non-cropped image coordinates of the center (xc,yc) = (892,875)

 

 

 

2014-07-16 Qads alignment

Image in previous alignment was not a shadow of a disk, but it was a gain map, so it is not correct to use it for alignment.

For further alignment we use data from exp=cxid6014:run=50

(tick) offset_corr/1-end.data

  0   -8  -1   6
  0    8   6   3
  0    0   0   0

Plots for alignment

Circle radius [pix] = 150

For non-cropped image coordinates of the center (xc,yc) = (892,875)

 

 

  • No labels