Content
Optical measurements
Comments from Gabriel Blaj:
DS1 was fixed, with only Q2 metrology updated. For Q0, Q1, Q3 please use previous metrology / corrections.
These metrology files should be used for all experiments AFTER May 15th (until next upgrade/repair).
- 2014-05-15-CSPAD-CXI-DS1-Metrology.xlsx
- 2014-05-15-CSPAD-CXI-DS1-Metrology.txt
- 2014-05-15-CSPAD-CXI-DS1-Metrology-corr.txt - is 2014-03-19-CSPAD-CXI-DS1-Metrology-corr.txt with new Q2 measurement
Quality check in X-Y plane
segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) Quad 0 segm: 0 20913 20911 35 33 43542 43545 -17 -14 0.04474 48305 48304 1 2 -3 segm: 1 20907 20911 6 10 43544 43547 -6 -3 0.01053 48304 48305 -1 -4 -3 segm: 2 20907 20908 -12 -13 43543 43545 -4 -2 -0.01579 48305 48300 5 1 -2 segm: 3 20914 20915 -21 -22 43552 43561 -19 -10 -0.02762 48313 48321 -8 1 -9 segm: 4 20909 20909 -39 -39 43597 43545 23 -29 -0.05128 48308 48347 -39 0 52 segm: 5 20904 20898 -24 -30 43557 43544 23 10 -0.03552 48309 48303 6 6 13 segm: 6 20934 20934 -256 -256 43578 43580 -127 -125 -0.33658 48344 48349 -5 0 -2 segm: 7 20912 20942 -33 -63 43550 43548 -23 -25 -0.06315 48315 48317 -2 30 2 Quad 1 segm: 0 20906 20905 68 67 43535 43541 -34 -28 0.08817 48298 48295 3 1 -6 segm: 1 20904 20902 -10 -12 43543 43537 6 0 -0.01448 48295 48299 -4 2 6 segm: 2 20907 20900 17 24 43538 43537 11 10 0.02632 48296 48295 1 -7 1 segm: 3 20902 20903 31 30 43539 43537 15 13 0.03948 48295 48296 -1 1 2 segm: 4 20906 20907 -1 0 43545 43545 -1 -1 0.00000 48302 48304 -2 -1 0 segm: 5 20909 20909 52 52 43542 43542 -27 -27 0.06843 48300 48303 -3 0 0 segm: 6 20907 20907 45 45 43542 43577 -14 21 0.05919 48300 48333 -33 0 -35 segm: 7 20908 20906 57 59 43541 43543 27 29 0.07632 48301 48301 0 -2 -2 Quad 2 segm: 0 20902 20908 -49 -43 43546 43542 26 22 -0.06053 48303 48300 3 -6 4 segm: 1 20906 20906 77 77 43544 43546 -40 -38 0.10132 48301 48305 -4 0 -2 segm: 2 20916 20867 29 78 43528 43608 -6 74 0.06970 48325 48310 15 -49 -80 segm: 3 20909 20917 9 1 43546 43544 7 5 0.00658 48309 48303 6 8 2 segm: 4 20911 20921 131 141 43544 43547 -63 -60 0.17895 48311 48305 6 -10 -3 segm: 5 20901 20935 74 108 43553 43547 -41 -47 0.11972 48313 48313 0 -34 6 segm: 6 20918 20924 -18 -24 43543 43570 -15 12 -0.02762 48328 48312 16 6 -27 segm: 7 20896 20918 30 8 43586 43548 -7 -45 0.02499 48292 48355 -63 22 38 Quad 3 segm: 0 20902 20899 72 69 43537 43538 -33 -32 0.09212 48295 48293 2 3 -1 segm: 1 20905 20909 48 52 43534 43545 -28 -17 0.06580 48300 48297 3 -4 -11 segm: 2 20907 20907 259 259 43539 43539 124 124 0.34083 48299 48299 0 0 0 segm: 3 20894 20901 16 9 43538 43537 10 9 0.01579 48296 48289 7 7 1 segm: 4 20904 20903 95 94 43541 43542 -46 -45 0.12369 48299 48299 0 1 -1 segm: 5 20904 20905 158 159 43541 43542 -77 -76 0.20791 48299 48300 -1 -1 -1 segm: 6 20903 20905 41 39 43540 43539 22 21 0.05264 48299 48295 4 2 1 segm: 7 20907 20904 11 14 43542 43543 5 6 0.01579 48300 48301 -1 -3 -1
Quality check in Z
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um) Quad 0 segm: 0 20912 43543 43543 20912 -6 9 -115 -100 1 -107 -15 -15 -107 1 -0.14080 0.00274 -15.007 segm: 1 20909 43545 43545 20909 -4 5 -96 -87 0 -91 -9 -9 -91 0 -0.11974 0.00000 -9.007 segm: 2 20907 43544 20907 43544 -48 -28 -84 -64 -38 -74 -20 -20 -38 -74 -0.10414 -0.09737 -20.006 segm: 3 20914 43556 20914 43556 -17 -4 -109 -96 -10 -102 -13 -13 -10 -102 -0.02740 -0.13418 -13.023 segm: 4 20909 43571 43571 20909 99 -53 162 10 23 86 152 152 86 23 0.11309 0.06303 151.804 segm: 5 20901 43550 43550 20901 -15 -72 -26 -83 -43 -54 57 57 -54 -43 -0.07104 -0.11788 57.012 segm: 6 20934 43579 20934 43579 2 5 -11 -8 3 -9 -3 -3 3 -9 0.00821 -0.01183 -3.001 segm: 7 20927 43549 20927 43549 -22 22 -27 17 0 -5 -44 -44 0 -5 0.00000 -0.00658 -44.030 Quad 1 segm: 0 20905 43538 43538 20905 76 76 24 24 76 24 0 0 24 76 0.03158 0.20830 -0.000 segm: 1 20903 43540 43540 20903 91 92 32 33 91 32 -1 -1 32 91 0.04211 0.24943 -1.013 segm: 2 20903 43537 20903 43537 18 17 57 56 17 56 1 1 17 56 0.04660 0.07370 0.993 segm: 3 20902 43538 20902 43538 27 36 45 54 31 49 -9 -9 31 49 0.08498 0.06448 -9.001 segm: 4 20906 43545 43545 20906 20 21 22 23 20 22 -1 -1 22 20 0.02895 0.05481 -0.999 segm: 5 20909 43542 43542 20909 1 28 93 120 14 106 -27 -27 106 14 0.13948 0.03836 -27.000 segm: 6 20907 43559 20907 43559 26 19 93 86 22 89 7 7 22 89 0.06029 0.11707 7.075 segm: 7 20907 43542 20907 43542 17 20 125 128 18 126 -3 -3 18 126 0.04933 0.16580 -2.996 Quad 2 segm: 0 20905 43544 43544 20905 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 1 20906 43545 43545 20906 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 2 20891 43568 20891 43568 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 3 20913 43545 20913 43545 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 4 20916 43545 43545 20916 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 5 20918 43550 43550 20918 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 6 20921 43556 20921 43556 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 segm: 7 20907 43567 20907 43567 0 0 0 0 0 0 0 0 0 0 0.00000 0.00000 0.000 Quad 3 segm: 0 20900 43537 43537 20900 97 102 -42 -37 99 -39 -5 -5 -39 99 -0.05132 0.27140 -5.015 segm: 1 20907 43539 43539 20907 128 111 -10 -27 119 -18 17 17 -18 119 -0.02369 0.32612 17.022 segm: 2 20907 43539 20907 43539 6 10 140 144 8 142 -4 -4 8 142 0.02192 0.18687 -4.000 segm: 3 20897 43537 20897 43537 -5 -1 115 119 -3 117 -4 -4 -3 117 -0.00823 0.15397 -4.004 segm: 4 20903 43541 43541 20903 19 36 -62 -45 27 -53 -17 -17 -53 27 -0.06974 0.07401 -17.002 segm: 5 20904 43541 43541 20904 48 47 -36 -37 47 -36 1 1 -36 47 -0.04737 0.12882 1.001 segm: 6 20904 43539 20904 43539 -7 5 222 234 -1 228 -12 -12 -1 228 -0.00274 0.30004 -12.006 segm: 7 20905 43542 20905 43542 -10 16 155 181 3 168 -26 -26 3 168 0.00822 0.22107 -25.995
center/0-end.data (pixel)
197.94 198.05 304.45 97.98 623.08 623.74 706.30 498.64 197.93 198.08 304.13 96.70 621.81 621.77 704.64 497.90 197.79 197.89 299.03 95.18 620.14 619.63 704.90 497.72 197.69 197.92 301.97 99.21 623.82 622.78 704.38 497.16 301.22 95.14 621.13 621.36 516.12 724.42 197.91 196.12 303.22 95.04 620.07 620.30 518.10 725.83 198.43 198.14 302.63 95.46 624.44 624.31 520.94 726.27 197.40 198.14 301.11 95.32 622.75 622.42 517.86 724.77 197.25 197.98 0.47 0.17 -0.25 0.17 0.65 -0.19 -0.14 -0.50 1.62 0.56 2.34 2.25 1.47 2.11 1.02 0.97 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.93 0.37 2.47 2.46 1.77 2.24 0.55 1.07
or the same in um
21757 21769 33464 10769 68489 68561 77637 54810 21757 21773 33430 10628 68349 68345 77454 54729 21741 21752 32869 10462 68166 68109 77482 54709 21730 21755 33193 10904 68570 68456 77425 54648 33110 10457 68275 68299 56732 79628 21754 21558 33329 10446 68158 68183 56949 79783 21811 21779 33265 10493 68638 68624 57261 79832 21698 21779 33098 10477 68452 68416 56923 79666 21681 21761 51 18 -28 18 71 -20 -15 -54 178 61 257 247 161 231 111 106 0 0 0 0 0 0 0 0 102 40 272 270 194 246 60 118
tilt/0-end.data (degree)
0.04474 0.01053 -0.01579 -0.02762 -0.05128 -0.03552 -0.33658 -0.06315 0.08817 -0.01448 0.02632 0.03948 0.00000 0.06843 0.05919 0.07632 -0.06053 0.10132 0.06970 0.00658 0.17895 0.11972 -0.02762 0.02499 0.09212 0.06580 0.34083 0.01579 0.12369 0.20791 0.05264 0.01579
geometry/0-end.data
# TITLE Geometry parameters of CSPAD # DATE_TIME 2014-10-16 10:24:57 PDT # METROLOGY /reg/neh/home/dubrovin/LCLS/CSPadMetrologyProc/2014-05-15-CSPAD-CXI-DS1-Metrology-corr.txt # AUTHOR dubrovin # EXPERIMENT MEC # DETECTOR CSPAD # CALIB_TYPE geometry # COMMENT:01 Table contains the list of geometry parameters for alignment of 2x1 sensors, quads, CSPAD, etc # COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame # PARAM:01 PARENT - name and version of the parent object # PARAM:02 PARENT_IND - index of the parent object # PARAM:03 OBJECT - name and version of the object # PARAM:04 OBJECT_IND - index of the new object # PARAM:05 X0 - x-coordinate [um] of the object origin in the parent frame # PARAM:06 Y0 - y-coordinate [um] of the object origin in the parent frame # PARAM:07 Z0 - z-coordinate [um] of the object origin in the parent frame # PARAM:08 ROT_Z - object design rotation angle [deg] around Z axis of the parent frame # PARAM:09 ROT_Y - object design rotation angle [deg] around Y axis of the parent frame # PARAM:10 ROT_X - object design rotation angle [deg] around X axis of the parent frame # PARAM:11 TILT_Z - object tilt angle [deg] around Z axis of the parent frame # PARAM:12 TILT_Y - object tilt angle [deg] around Y axis of the parent frame # PARAM:13 TILT_X - object tilt angle [deg] around X axis of the parent frame # HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X QUAD:V1 0 SENS2X1:V1 0 21757 33110 51 0 0 0 0.04474 -0.14079 -0.00274 QUAD:V1 0 SENS2X1:V1 1 21769 10457 18 0 0 0 0.01053 -0.11974 0.00000 QUAD:V1 0 SENS2X1:V1 2 33464 68275 -28 270 0 0 -0.01645 0.10414 0.09737 QUAD:V1 0 SENS2X1:V1 3 10769 68299 18 270 0 0 -0.02828 0.02740 0.13418 QUAD:V1 0 SENS2X1:V1 4 68489 56732 71 180 0 0 -0.05128 -0.11309 0.06303 QUAD:V1 0 SENS2X1:V1 5 68561 79628 -20 180 0 0 -0.03552 0.07104 -0.11788 QUAD:V1 0 SENS2X1:V1 6 77637 21754 -15 270 0 0 -0.33657 -0.00821 0.01183 QUAD:V1 0 SENS2X1:V1 7 54810 21558 -54 270 0 0 -0.06315 0.00000 0.00658 QUAD:V1 1 SENS2X1:V1 0 21757 33329 178 0 0 0 0.08883 0.03158 -0.20830 QUAD:V1 1 SENS2X1:V1 1 21773 10446 61 0 0 0 -0.01448 0.04211 -0.24943 QUAD:V1 1 SENS2X1:V1 2 33430 68158 257 270 0 0 0.02698 -0.04660 -0.07370 QUAD:V1 1 SENS2X1:V1 3 10628 68183 247 270 0 0 0.04014 -0.08498 -0.06448 QUAD:V1 1 SENS2X1:V1 4 68349 56949 161 180 0 0 -0.00066 -0.02895 0.05481 QUAD:V1 1 SENS2X1:V1 5 68345 79783 231 180 0 0 0.06843 -0.13948 0.03836 QUAD:V1 1 SENS2X1:V1 6 77454 21811 111 270 0 0 0.05919 -0.06029 -0.11707 QUAD:V1 1 SENS2X1:V1 7 54729 21779 106 270 0 0 0.07632 -0.04933 -0.16580 QUAD:V1 2 SENS2X1:V1 0 21741 33265 0 0 0 0 -0.06053 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 1 21752 10493 0 0 0 0 0.10132 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 2 32869 68638 0 270 0 0 0.07036 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 3 10462 68624 0 270 0 0 0.00658 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 4 68166 57261 0 180 0 0 0.17894 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 5 68109 79832 0 180 0 0 0.11972 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 6 77482 21698 0 270 0 0 -0.02762 0.00000 0.00000 QUAD:V1 2 SENS2X1:V1 7 54709 21779 0 270 0 0 0.02499 0.00000 0.00000 QUAD:V1 3 SENS2X1:V1 0 21730 33098 102 0 0 0 0.09278 -0.05132 -0.27140 QUAD:V1 3 SENS2X1:V1 1 21755 10477 40 0 0 0 0.06580 -0.02369 -0.32612 QUAD:V1 3 SENS2X1:V1 2 33193 68452 272 270 0 0 0.34083 -0.02192 -0.18687 QUAD:V1 3 SENS2X1:V1 3 10904 68416 270 270 0 0 0.01645 0.00823 -0.15397 QUAD:V1 3 SENS2X1:V1 4 68570 56923 194 180 0 0 0.12435 0.06974 0.07401 QUAD:V1 3 SENS2X1:V1 5 68456 79666 246 180 0 0 0.20857 0.04737 0.12882 QUAD:V1 3 SENS2X1:V1 6 77425 21681 60 270 0 0 0.05264 0.00274 -0.30004 QUAD:V1 3 SENS2X1:V1 7 54648 21761 118 270 0 0 0.01645 -0.00822 -0.22107 CSPAD:V1 0 QUAD:V1 0 -4500 -2700 0 90 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 1 -3500 5800 0 0 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 2 5500 4500 0 270 0 0 0.00000 0.00000 0.00000 CSPAD:V1 0 QUAD:V1 3 4500 -4500 0 180 0 0 0.00000 0.00000 0.00000
2014-06-05 Alignment of quads
hdf5 file with CSPAD array was kindly presented by Anton Barty for
exp=cxid2714:run=23
offset_corr/0-end.data
0 1 2 5 0 1 3 4 0 0 0 0
Overview
Content Tools