Alignment of quads

The geometry alignment parameters for cxi64813 have been copied from one of the earlier CXI experiments.
For test we use cxi64813-r0058.h5 with PyCSPadImage/src/Alignment.py
To adjust quads the files were changed to:
offset

  0    0  820  820
  0  820  820    0
  0    0    0    0

offset_corr

  0   0    5   2
  0   5    4   4
  0   0    0   0

marg_gap_shift

15   40   0  37
15   40   0  37
 0    0   0   0

Test image

Averaged 10 and 100 events:

Test image for cxi80410-r1458

Averaged over 500 events, pedestals (r1453) subtracted: