The geometry alignment parameters for cxi49012 have been copied from cxi80410-r1150.
For test we use cxi49012 data from runs 8(dark) 20(rings), written on 2012-02-02.
To adjust quads the files were changed as follows:

offset

  0    0  820  820
  0  820  820    0
  0    0    0    0

All offsets 834 were changed to 820 in order to get more space for positive gaps at minimal other parameters.
offset_corr

 1  0 -4 -3
 0  1  1  5
 0  0  0  0

marg_gap_shift

15   40   4  32
15   40   4  32
 0    0   0   0

The offset_corr were adjusted for gapx=gapy=0, when quad's 2x1 touch to each other without overlap. Then, the gaps were increased to 4:

Images for gap=0

Image for gap=2, and gap=4