Optical measurement
2012-02-26 XPP CSPad metrology files from Chris Kenney:
2012-02-26-CSPAD-XPP-Configuration.pptx
2012-02-26-CSPAD-XPP-Metrology.xlsx
identically converted to the text format:
2012-02-26-CSPAD-XPP-Metrology.txt
Chris has changed the order of measurement: in stead of individual quads entire detector sensor positions are meaasured in changed order:
The new script
/reg/neh/home/dubrovin/LCLS/test_optical/convertOpticalToStandard.py
converts the new measurement to the standard file format:
2012-02-26-CSPAD-XPP-Metrology-standard.txt
Procedure and table of results
We calculate
S1 - 1st short side of 2x1
S2 - 2nd short side of 2x1
L1 - 1st long side of 2x1
L2 - 2nd long side of 2x1
dS and dL are the deviations of the 1st and 2nd corner along the short and long sides, respectively. The sign of all dS are chosen in order to provide correct sign for the tilt angle (the same direction for all 2x1 sensors).
Everything, excluding <dS/L> and angle(deg), are in micrometers.
pair: S1 S2 dS1 dS2 L1 L2 dL1 dL2 <dS/L> angle(deg) Quad 0 pair: 0 20903 20913 -321 -311 43536 43528 154 146 -0.00726 -0.41591 pair: 1 20937 20904 -255 -288 43532 43534 130 132 -0.00624 -0.35733 pair: 2 20904 20905 -64 -65 43535 43536 -30 -29 -0.00148 -0.08489 pair: 3 20903 20902 -4 -3 43541 43539 -2 -4 -0.00008 -0.00461 pair: 4 20909 20893 -313 -329 43541 43534 153 146 -0.00737 -0.42244 pair: 5 20903 20902 -9 -10 43543 43541 3 1 -0.00022 -0.01250 pair: 6 20905 20898 83 90 43541 43548 38 45 0.00199 0.11382 pair: 7 20900 20901 139 138 43539 43541 66 68 0.00318 0.18226 Quad 1 pair: 0 20904 20903 -576 -577 43543 43536 280 273 -0.01324 -0.75864 pair: 1 20904 20902 -409 -411 43544 43540 201 197 -0.00942 -0.53951 pair: 2 20906 20895 -104 -93 43537 43536 -43 -44 -0.00226 -0.12963 pair: 3 20902 20909 -159 -166 43539 43539 -76 -76 -0.00373 -0.21384 pair: 4 20898 20904 -539 -533 43542 43538 262 258 -0.01231 -0.70534 pair: 5 20909 20689 -192 -412 43301 43542 192 433 -0.00695 -0.39810 pair: 6 20904 20903 -90 -89 43538 43543 -40 -35 -0.00206 -0.11777 pair: 7 20901 20903 -20 -22 43538 43538 -9 -9 -0.00048 -0.02764 Quad 2 pair: 0 20899 20916 -369 -352 43537 43537 169 169 -0.00828 -0.47443 pair: 1 20919 20896 -287 -310 43539 43545 136 142 -0.00686 -0.39279 pair: 2 20906 20909 -222 -225 43542 43542 -108 -108 -0.00513 -0.29410 pair: 3 20906 20904 -185 -183 43542 43542 -83 -83 -0.00423 -0.24212 pair: 4 20911 20906 -331 -336 43537 43536 160 159 -0.00766 -0.43890 pair: 5 20897 20907 -294 -284 43536 43538 131 133 -0.00664 -0.38033 pair: 6 20895 20912 -55 -72 43535 43539 -28 -24 -0.00146 -0.08357 pair: 7 20903 20903 -178 -178 43531 43543 -98 -86 -0.00409 -0.23425 Quad 3 pair: 0 20897 20904 -188 -181 43539 43533 98 92 -0.00424 -0.24281 pair: 1 20902 20907 3 8 43543 43536 2 -5 0.00013 0.00724 pair: 2 20906 20904 -60 -58 43538 43537 -26 -27 -0.00136 -0.07764 pair: 3 20907 20906 57 58 43535 43540 27 32 0.00132 0.07567 pair: 4 20909 20896 -137 -150 43530 43543 61 74 -0.00330 -0.18885 pair: 5 20904 20904 35 35 43541 43544 -15 -12 0.00080 0.04606 pair: 6 20905 20907 -143 -145 43540 43539 -64 -65 -0.00331 -0.18950 pair: 7 20905 20909 -150 -154 43537 43539 -68 -66 -0.00349 -0.20003
- All ds1, dl1 are well consistent (within +/-10um) with ds1, dl1, respectively
- Angles of 2x1 are in the range from -0.76 to +0.18 degree.
- All tilts are measured for entire detector. No need to apply any tilt for quads.
Configuration parameters
center/0-end.data
200.40 198.61 311.79 99.25 629.11 629.68 712.05 500.23 201.15 198.98 308.78 95.96 625.01 626.79 712.68 499.87 200.62 198.68 311.91 99.60 627.98 629.45 713.34 500.04 198.70 198.06 309.46 96.33 627.20 627.47 710.18 497.82 305.89 94.00 624.26 625.19 512.45 725.32 196.23 197.12 306.38 93.22 623.61 624.40 510.94 724.06 195.52 195.79 306.15 93.80 623.01 624.23 511.13 723.50 195.84 197.00 307.67 95.10 624.72 624.64 510.01 722.80 192.77 193.90 -0.27 0.03 -0.64 -0.42 -0.93 -1.16 -0.66 -0.41 0.22 0.45 -0.16 -0.39 0.44 0.03 0.83 -0.27 0.90 0.65 1.26 1.22 1.27 1.52 0.73 0.88 0.32 0.23 0.31 0.42 -0.08 0.04 -0.36 -0.09
tilt/0-end.data
-0.41591 -0.35733 -0.08489 -0.00461 -0.42244 -0.01250 0.11382 0.18226 -0.75864 -0.53951 -0.12963 -0.21384 -0.70534 -0.39810 -0.11777 -0.02764 -0.47443 -0.39279 -0.29410 -0.24212 -0.43890 -0.38033 -0.08357 -0.23425 -0.24281 0.00724 -0.07764 0.07567 -0.18885 0.04606 -0.18950 -0.20003
offset/0-end.data
0 0 820 820 0 820 820 0 0 0 0 0
offset_corr/0-end.data
0 3 3 0 0 0 0 0 0 0 0 0
marg_gap_shift/0-end.data
15 40 2 25 15 40 2 25 0 0 0 0
geometry/0-end.data
# TITLE Geometry parameters of CSPAD-XPP # DATE_TIME 2014-10-03 11:59:59 PDT # METROLOGY /reg/neh/home1/dubrovin/LCLS/CSPadMetrologyProc/2012-02-26-CSPAD-XPP-Metrology.txt # AUTHOR dubrovin # EXPERIMENT Any # DETECTOR CSPAD-XPP # CALIB_TYPE geometry # COMMENT:01 Table contains the list of geometry parameters for alignment of 2x1 sensors, quads, CSPAD, etc # COMMENT:02 All translation and rotation pars of the object are defined w.r.t. parent object Cartesian frame # PARAM:01 PARENT - name and version of the parent object # PARAM:02 PARENT_IND - index of the parent object # PARAM:03 OBJECT - name and version of the object # PARAM:04 OBJECT_IND - index of the new object # PARAM:05 X0 - x-coordinate [um] of the object origin in the parent frame # PARAM:06 Y0 - y-coordinate [um] of the object origin in the parent frame # PARAM:07 Z0 - z-coordinate [um] of the object origin in the parent frame # PARAM:08 ROT_Z - object design rotation angle [deg] around Z axis of the parent frame # PARAM:09 ROT_Y - object design rotation angle [deg] around Y axis of the parent frame # PARAM:10 ROT_X - object design rotation angle [deg] around X axis of the parent frame # PARAM:11 TILT_Z - object tilt angle [deg] around Z axis of the parent frame # PARAM:12 TILT_Y - object tilt angle [deg] around Y axis of the parent frame # PARAM:13 TILT_X - object tilt angle [deg] around X axis of the parent frame # HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X CSPAD:V2 0 SENS2X1:V1 0 52033 113005 -29 90 0 0 -0.41590 0.08769 -0.07239 CSPAD:V2 0 SENS2X1:V1 1 75324 112808 3 90 0 0 -0.35733 0.00548 -0.04475 CSPAD:V2 0 SENS2X1:V1 2 17038 125249 -70 0 0 0 -0.08489 0.05528 0.09867 CSPAD:V2 0 SENS2X1:V1 3 16936 101886 -46 0 0 0 -0.00461 0.07632 -0.00274 CSPAD:V2 0 SENS2X1:V1 4 29328 160129 -102 270 0 0 -0.42243 -0.02741 0.08554 CSPAD:V2 0 SENS2X1:V1 5 5930 160191 -127 270 0 0 -0.01250 -0.10691 0.08553 CSPAD:V2 0 SENS2X1:V1 6 64087 169245 -73 0 0 0 0.11382 0.05658 0.12336 CSPAD:V2 0 SENS2X1:V1 7 63990 145961 -44 0 0 0 0.18226 0.05922 0.04386 CSPAD:V2 0 SENS2X1:V1 8 108743 129761 24 0 0 0 -0.75860 0.06448 0.10964 CSPAD:V2 0 SENS2X1:V1 9 108504 106332 49 0 0 0 -0.53949 0.06843 0.06304 CSPAD:V2 0 SENS2X1:V1 10 120574 164631 -18 270 0 0 -0.12963 -0.09047 0.11844 CSPAD:V2 0 SENS2X1:V1 11 97181 164719 -43 270 0 0 -0.21384 -0.11511 0.12370 CSPAD:V2 0 SENS2X1:V1 12 155334 152248 48 180 0 0 -0.70531 -0.08290 -0.07401 CSPAD:V2 0 SENS2X1:V1 13 155529 175673 2 180 0 0 -0.39849 -0.05674 -0.17355 CSPAD:V2 0 SENS2X1:V1 14 164970 117576 91 270 0 0 -0.11777 -0.01096 0.07764 CSPAD:V2 0 SENS2X1:V1 15 141579 117606 -30 270 0 0 -0.02764 -0.29604 0.05001 CSPAD:V2 0 SENS2X1:V1 16 125593 73085 99 270 0 0 -0.47442 -0.05481 0.03027 CSPAD:V2 0 SENS2X1:V1 17 102251 73298 71 270 0 0 -0.39278 -0.12606 0.02105 CSPAD:V2 0 SENS2X1:V1 18 160422 60852 138 180 0 0 -0.29410 -0.06185 -0.03015 CSPAD:V2 0 SENS2X1:V1 19 160556 84189 133 180 0 0 -0.24212 -0.05790 -0.06030 CSPAD:V2 0 SENS2X1:V1 20 148124 26109 139 90 0 0 -0.43889 0.08495 -0.02895 CSPAD:V2 0 SENS2X1:V1 21 171468 25948 167 90 0 0 -0.38033 0.02193 -0.01184 CSPAD:V2 0 SENS2X1:V1 22 113467 16726 80 180 0 0 -0.08357 -0.07370 0.02467 CSPAD:V2 0 SENS2X1:V1 23 113595 40172 97 180 0 0 -0.23425 -0.04738 -0.04112 CSPAD:V2 0 SENS2X1:V1 24 68896 56100 35 180 0 0 -0.24281 -0.06449 0.01371 CSPAD:V2 0 SENS2X1:V1 25 68966 79466 25 180 0 0 0.00724 -0.08159 -0.06304 CSPAD:V2 0 SENS2X1:V1 26 56721 21251 34 90 0 0 -0.07764 0.09593 -0.00263 CSPAD:V2 0 SENS2X1:V1 27 80148 21259 46 90 0 0 0.07567 0.07400 -0.02106 CSPAD:V2 0 SENS2X1:V1 28 21795 33860 -8 0 0 0 -0.18885 0.03553 0.05208 CSPAD:V2 0 SENS2X1:V1 29 21765 10469 4 0 0 0 0.04606 0.02632 0.01645 CSPAD:V2 0 SENS2X1:V1 30 12674 68730 -40 90 0 0 -0.18950 0.07674 -0.03290 CSPAD:V2 0 SENS2X1:V1 31 36016 68606 -10 90 0 0 -0.20003 0.03289 -0.02500
Alignment check
Use xppcom10-r1437 for wire image and r1442 for dark image: