To reinitialize motors when powering back on:

Waveplates (Newport XPS 4 motors)

  • Go to Laser>Laser (Systems)>Other>Newport XPS_4
  • Waveplates GH/IJ


  • Click “initialize all”
  • Wait for each motor to refresh with “initialize / “Reference”
  • At the top of the page click “Initialize all”
  • At the top of the page click “Reference all”


Waveplates AB/EF with MForce connection

  • Go to Laser>Laser Systems>NS-Laser>Polarizer


  • Click on the  “expert” button in the epics window       
  • Click on the “Clear Power-Up” button


SHG  (Newport XPS3 motors)

    • Go to Laser>Laser Systems>Other>Newport XPS>Newport XPS_3 (TimeTool)
    • SHG AB/EF/GH/IJ
    • Perform the same procedure as listed above for Newport XPS 4