Beam Alignemnt
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
DSHM | OTRS:LI20:3075:MOTR | -37200/-500 | in/out | ||
FrontView | XPS:LI20:MC05 | Target Mount Horizontal | M2 | -120/-150 | in/out |
FrontView | XPS:LI20:MC05 | Target Mount Vertical | M1 | -20/-20 | no change |
Ionizer/Shadowgraphy
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
Afterglow injector | XPS:LI20:MC01 | E12 Afterglow injector | M1 | +28 mm/ +4 mm/ -46 mm | Collision closest to PB/ Focus / collision furthest from PB |
Shadowgraphy camera stage | XPS:LI20:MC02 | E14 Shadowgraphy Imaging | M1 | Moves the camera between MO and LFOV imaging | |
Microscope objective | MOC-LI20-MC08 | PB28 MO Vertical Mover | CH3 | Picomotor, controls the MO |
Laser room
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
HeNe beam block | XPS:LA20:LS24 | Transport HeNe block | M3 | 10 mm/ 8 mm | in/out |
CW beam block | XPS:LA20:LS24 | CW laser block | M5 | 0 mm/ 25 mm | in/out |
Main beam block
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
Main beam block | XPS:LI20:MC03 | C9 Compressor beam block | M3 | 299 mm/ 0mm | in/out |
E300
For sending laser to foil, equivalent of minimum transmission through main attenuator + RP attenuator in is needed.
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
USHM | XPS:LI20:MC03 | PB10 M2 Removal Stage | M1 |
0.5/42.5 | in/out (E320 position) |
Axicon Vertical | XPS:LI20:MC04 | PB2 Lens Vertical | M2 | 0.9 mm | Axicon beam center overlaps with beam center |
Axicon Horizontal | XPS:LI20:MC04 | PB3 Lens Horizontal | M3 | 10 mm | Axicon beam center overlaps with beam center |
Axicon z | XPS:LI20:MC04 | PB1 Lens Z Position Stage | M1 | 0 | Position for Axicon |
E305
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
Axilens z | XPS:LI20:MC04 | PB1 Lens Z Position Stage | M1 | 20 | Position for Axilens |
Axilens Vertical | XPS:LI20:MC04 | PB2 Lens Vertical | M2 | 0/100 in/out | Axicon beam center overlaps with beam center |
Axilens Horizontal | XPS:LI20:MC04 | PB3 Lens Horizontal | M3 |
8.57 | Axicon beam center overlaps with beam center |
USHM | XPS:LI20:MC03 | PB10 M2 Removal Stage | M1 |
3.5 mm / 44 mm 0.5/42.5 | in/out (E320 position) |
EOS
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
EOS removal | XPS:LI20:MC04 | PB4 EOS Assembly Removal | M4 | 10.7 mm/ 50mm | in/out |
Target assembly
Desc | XPS Controller | Motor name | Channel | Position | Desc. position |
---|---|---|---|---|---|
Target Assembly Vertical | XPS:LI20:MC05 | PB7 Target Mount Vertical | M1 | 80 | Clear in combination with horizontal |
Target Assembly Horizontal | XPS:LI20:MC05 | PB8 Target Mount Horizontal | M2 | -150 | Clear in combination with vertical |
E320 assembly
Attention: the alignment target can crash with multiple items. Please contact E-320 if you don't know the procedures.
Desc | XPS Controller | Motor name | Channel | Position | Desc. position | Last update |
---|---|---|---|---|---|---|
USHM | XPS:LI20:MC03 | PB10 M2 Removal Stage | MC03, M1 | OAPs in:
laser vector alignment: 0.5 (cross-check with 305 value above) | E-320 position: mirror on the right side of the e-beam, steering laser onto 1st OAP | April 30, 2024 |
Alignment Target Vertical UP/DOWN | XPS:LI20:MC10 | PB11 SFQED Target Vertical | MC10, M1 | 0-9048.35 | highest position / lowest position / out position when OAP is IN / YAG in position ATTENTION: collisions possible, follow exact procedure, check target position before moving OAPs / MO tower DON'T GO LARGER THAN 90 EVER http://physics-elog.slac.stanford.edu/facetelog/show.jsp?dir=/2024/17/24.04&pos=2024-04-24T23:48:34 | April 24, 2024 |
MO Tower Mover IN/OUT | XPS:LI20:MC10 | PB12 SFQED Tower Mover | MC10, M2 | -95/0 | IN position to look at focal spot.
| December 14, 2022 |
OAP Assembly IN/OUT | XPS:LI20:MC10 | PB16 SFQED Assembly Removal | MC10, | IN position to focus main beam. Clear (retracted), also home positions, needs MO Tower to move out first, also needs Alignment target to move upwards first | April 30, 2024 | |
Elliptical mirror/M3 IN/OUT | XPS:LI20:MC10 | PB17 Elliptical mirror mover | MC10, M4 | Positive limit switch/negative limit switch + 7 | Out position. Configuration is 'SetCurrentPositionAsHome',so need to drive all the way to positive limit switch to ensure driven out. In position. Drive all the way to negative end for IN position. | |
PB18 Focus Imaging Vertical | XPS:LI20:MC10 | PB18 Focus Imaging Vertical | MC10, M5 | 0.427 | Home at current position (0). Motor should not be touched or moved, should be clear in any case. | |
Foils vertical mover | XPS:LI20:MC05 | MC05, M1 | needed only for beam-vector alignment (M1/M2) | November 8, 2023 | ||
Foils horizontal mover | XPS:LI20:MC05 | MC05, M2 | needed only for beam-vector alignment (M1/M2) | November 8, 2023 |
December 14, 2022: http://physics-elog.slac.stanford.edu/facetelog/show.jsp?dir=/2022/50/14.12&pos=2022-12-14T23:59:00
April 24, 2024 E320 OAP swap and new motor positions: http://physics-elog.slac.stanford.edu/facetelog/show.jsp?dir=/2024/17/24.04&pos=2024-04-24T23:48:34
Safely retracting E-320 assembly and setting up to clear e-beam/laser will be:
- Drive PB12 SFQED Tower Mover to 0 (home)
- Drive PB11 SFQED Target Vertical to 0 (home)
- Drive PB16 SFQED Assembly Removal to 0 (home)
- Drive PB11 SFQED Target Vertical to 90
- Drive PB17 Elliptical mirror mover to most positive end
Note: TRA motors on MC10 currently do not work in EPICS (position is shown correctly). Drive through XPS web interface.