SOP: Vacuum and Flow Operation (Thanks to Mike and Thomas)
Transfer the SOP from the above link:
Notes
Unlike in XPP, the manifold does not need a pressure relief valve in CXI. The CXI chamber has a pressure relief burst disk. Due to the volume of the liquid sample reservoir, they cannot be attached to adjacent valcles. Therefore, every other valve (say, 4 and 6) should attach to a liquid sample, and the valves in between should attach to lecture bottles of gases. All tubing is ½” OD, until it adapts to 3/16” to the diffractometer.
Loading a liquid sample
- Separate the reservoir line from the rest of the manifold at the joint just above valve 4 or 6, whichever you use. (Requires size 1-1/8” and 15/16”wrenches.)
- Bring sample reservoir to a fume hood to fill it.
- Open the sample reservoir by separating the lid at the ISO flange. (Requires 1/2” wrench)
- Add liquid sample to the reservoir.
- Reattach lid of reservoir and make sure valve 4/6 is closed (to avoid spreading odor.)
- Reattach reservoir to manifold line. (Requires size 1-1/8” and 15/16” wrenches.)
Initialize vacuum
- Ensure that any new connections are properly tightened.
- Close all valves in the manifold.
- Open the gate valve between the backing pump and the turbo pump. (Not pictured)
- Turn on the backing pump.
- Wait for backing pump to stabilize.
- Open valve 3, then valves 2 & 8.
- Wait for backing pump to stabilize.
- Slowly open valve 1. → Pressure should be below the Baratron’s range.
- Turn on chiller for the turbo pump.
- Turn on the turbo pump.
- Wait for turbo pump to achieve standby mode (operation at less than 10W.)
Flowing a liquid sample
- Re-establish vacuum:
- Temporarily close gate valve between the turbo pump and the backing pump.
- Open valve 3, then 2 and 8 to evacuate air from manifold.
- Close valves 2 and 8.
- Open valve 4/6 to evacuate head space in the reservoir.
- Close valve 4/6.
- Close valve 3.
- Open gate valve between the turbo pump and the backing pump.
- Adjust flow rate up to desired pressure:
- Open valve 4/6.
- Open valve 8.
- Monitor Baratron pressure while slowly opening valve 1 (needle valve.)
- Attempt to approach desired pressure from lower pressure (try not to overshoot.)
Flowing a gas sample
- Close all valves.
- If the manifold has not yet been cleared of the last sample:
- Temporarily close the gate valve between the two pumps.
- Open valve 3, then 2 and 8 to evacuate manifold.
- Open valve 1 (needle valve.)
- Open valve 5 or 7, whichever you are using, to evacuate gas line.
- Close valve 5/7.
- Close valve 3, then 2 and 8.
- Reopen gate valve.
- Slowly adjust regulator to achieve a pressure of a few psi (unreadably low on regulator.)
- Open valve 5 or 7.
- Open valve 8.
- Slowly begin to open valve 1 (needle valve) to achieve desired manifold pressure by monitoring Baratron reading.
Manifold Images