Changing the quad settings
facet home → Profile Monitor → LI20 → SPEC LINE GUI
- Leave big "Dipole" switch in "Without" setting (dipole energy is not adjusted)
- Set "Z Object" to the position where the beam should be imaged, e.g., the center of the PB
- Set "Z Image" to the position of the screen, e.g., z=2015.26 for DTOTR
- M12, M34: transfer-matrix elements (set both to 0 for re-imaging)
- Energy: determines where the focus is
- Press "Calculate and Trim"
Changing the dipole settings
MC01 - Ch6 0 (DRZ in) 55 (everything out) 145 (CsI in) → before dipole scan move CsI out
e-beam should never be on CsI
LFOV with DRZ in for low-energy electrons
LI20 Experimenter beckhoff Channel 21 LFOV filter lipper (out: see low-energy tail, main beam saturated; in: light attenuated, main beam is not saturated)