You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 8 Next »

New Cameras

New cameras are in a plastic bin in the B244 back room.

Procedure

Example: Replacing IPOTR1, IPOTR2, and DSOTR

  1. Align HeNe to beam marks on FrontView, IPOTR1, IPOTR2, and DSOTR.
    1. This can be done with IP area vented or under vacuum. Note that you will have to realign HeNe going from vacuum to vented and vice-versa.
  2. Replace burned cameras with new cameras.
    1. IPOTR1/2 are Manta G-095B cameras. They can be replaced with Mako or Manta G-125B cameras. If Mako, you might need to adjust how the camera is mounted.
    2. Label cameras and note new serial numbers in the eLog.
    3. After replacing the camera, reboot to connect. If the camera does not connect, an IOC reboot might be necessary. Rebooting the IOC should connect the new camera.
  3. Verify alignment and focus.
    1. The HeNe should be in the center of the camera image.
    2. The “mark” position on the OTR foils should be clear and in focus. You may need to use a flashlight for illumination.

Reset Reboot Count

After the camera has been replaced, reset the reboot count to 0:

[fphysics@facet-srv01 ~/shperez ]$ caput CAMR:LI20:102:REBOOTCOUNT 0


Also double check that the serial number has updated:

[fphysics@facet-srv01 ~/shperez ]$ caget CAMR:LI20:102:SerialNumber_RBV

CAMR:LI20:102:SerialNumber_RBV 50-0503525584
  • No labels