PLC setup from ECS is here

Camera: Alvium 1800 U-1236m, 04K7X

3 min to move it IN

magnification

  • x2
  • x5
  • x10
  • x20

General comments on pointing

When looking from downstream to upstream TCC, and having the camera flipped in Y only, going - moves the pin to the right (since it moves the pin to the left on Q1).

Homing and presets

DescriptionStateSoft Limit switchComments
Filter 125: ND6-

2 on the MPI body, home forward →

0: ND4-1 on the MPI body, home forward →
-25: No filter-0 on the MPI body, home forward →
Filter 225: ND3-2 on the MPI body, home forward →
0: ND1.5-1 on the MPI body, home forward →
-25: No filter-0 on the MPI body, home forward →
Filter 325: ND1-2 on the MPI body, home forward →
0: color filter-1 on the MPI body, home forward →
-25: No filter-0 on the MPI body, home forward →
Shutter14: CLOSED14Home forward →
-10: OPEN-10Home forward →
X13.624: IN-

Home to negative limit switch

Moving + moves the pin to the left of the screen (assuming the camera is flipped in Y only)

Y17.: IN-

Home to negative limit switch

Moving + moves the pin up in the screen (assuming the camera is flipped in Y only)


97: OUT-
Z92.030: x20-Home to negative limit switch

: x10

-

: x5

-

:x2

-

5: OUT

-
  • No labels