CERN SR1 Setup
Requirements
| Number of Local Supports to accommodate | One Stave + One Ring |
| Maximum Dew Point | -60 C |
| Light tightness | any? |
| Electrical feed-through to connect two LS at the same time | Two independent or one with internal split box |
| Cooling feed-through for cold CO2 (-35C) hoses | One in, one out |
| Environmental sensors w/ interlock | Humidity, Temperature, pressure |
| LS Interlocks | Interlock |
Instruments
| Description | Scope | Make | |
|---|---|---|---|
| Infrared camera | Thermal QC | FLIR | We expect to measure the sensor temperature in the range of -35 to RT. The operating environment is ~ -10 C |
| Radioactive Source | Electrical QC | SLAC | Define type and Intensity. We need envelope of containment box and requirement for the handling and operation |
| Laser Interferometer | Mechanical QC | STIL | Optical Pen - STIL OP10000 + CCS PRIMA http://point.stil-sensors.com/?lang=EN |
QC test (Validation and or Verification)...this link eventually to the QA/QC documents and the requirements document
| Thermal performances | Thermal Figure of Merit; Temperature uniformity on Module or Heater | |
| Thermo-mechanical Deformation | Deformation of the LS under the effect of the cooing | |
| Electrical I | Power Distribution | |
| Electrical II | data transmission | |
| Electrical III |
|