Identification Details and Hardware Changelog

FirmwareCarrierID0CarrierID1DigitalCardID0DigitalCardID1AnalogCardID0AnalogCardID1Date
3925999620100024832140768634981794137088041030657124024542730067108888April 2022

0xea020004 = 3925999620


Performance Measures

Pedestals and Noise

For Pedestals, method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad (non-conforming pixels aka NCP).

exp:run (mm/dd/yy)

Temp (oC)Humidity (%)Mean Pedestal Value (ADU) Number of Ped NCPPedestal GraphsNoise (ADU)Number of Noise NCPNoise Graphs
detdaq21, r69 (08/01/24)8.824.134444 ± 283375

Histogram/Pixel Map

32115

Histogram/Pixel Map 

Bad Pixel Map

Map all pixels with a non-zero bad status bit, which is defined here

As of Aug 1, 2024 – 438 bad pixels

             

Configuration Settings



Experiment History:

RunExperimentHutchShiftsDate Deployed
22X-10043XCS4

 


X-10004XCS5

 


L-10267XPP5

 


L-10057XPP1

23L-10366 InoueXPP5

 



Gain Calibration 

Cu fluorescence, detector ~5cm from target, mini-x parameters: 15kvP, 200uA

in experiment detdaq21

typerun number
dark69
low flux70, 71, 72 (each ~10min)