Recovering from errors:  restarting the DAQ

Use the "restartdaq <-w>" command. If you only want to stop the DAQ, call "stopdaq".

The command "wheredaq" will tell if you an instance of the DAQ is currently running (e.g. on the hutch-machine when you are in the control room).


Recovering from errors:  restarting AMI (the client)

Use the "startami" command. This will start a second client if you are not on the main DAQ machine and it will restart the DAQ AMI client if run on the main DAQ machine (after asking for confirmation, so you will need to use a terminal). Should the server side by unhappy, you will need to restart the DAQ.

Troubleshooting the DAQ

Ami does not work: the DAQ is fine after a restart, but I see no updates/detectors in ami

Make sure that all dss nodes are selected. If you needed to take a node out due to problems, you need to edit the <hutch>.cnf file. You can use "serverStat <ip>" to get the name for the node that causes a problem and then edit the list of dss_nodes to exclude this node. If the problematic node is the last one, you might see that you have to reselect the Bld upon a restart. This means that the DAQ will make you allocate twice (the first time it'll fail with a complaint about a Bld change).

A second cause for this problem can be that the calib-dir is not mounted on the mon-nodes anymore. ami will look for the pedestal files here, so it will take a long time until int times out and will start, but w/o the detectors. You can check if that is the case by looking at an ami log file (e.g. for ami0-0) or ssh to a moo-node in use and  check if you can see the pedestals files in /reg/d/psdm/<hutch>/<expname>/calib/<dettype>/.... At this point, please call your on-call ECS person who will escalate it or work around it.

The DAQ shows "NO CONNECT" in procStat: 

use "serverStat <DAQ device alias> <command>". "cycle" will power cycle the node with some time between off/on in the script. It will tie up the terminal, so if you have to deal with several nodes, you can also call "serverStat <DAQ device alias> off" and "serverStat <DAQ device alias> on" explicitly. Remember to wait a few second between off/on. After the script returns from turning the node(s) on, continue to run "serverStat <ip/node name>" until both pings work. If you can ssh into the node(s), you can restart the DAQ.

The DAQ shows an error message, requesting a restart,  indicating a given IP as culprit

Use "serverStat <ip>" to check if both interfaces of the node in question are up. This script will also tell you which machine has the issue.

One/both of the pings fails: use "serverStat <ip/node name> cycle" to power cycle the machine. After the script returns, continue to run "serverStat <ip/node name>" until both pings work. If you can ssh into the node, you can restart the DAQ.

Otherwise: decide if you'd rather restart the DAQ and hope for the best. Power cycling a machine takes a few minutes.

The IP is a dss-node: here you have an additional option: you can edit the <hutch>.cnf file to take out the node: look for "dss_nodes = [....]" and take out the problematic node. 

Depending on the data rate, you can run with 2 or 3 nodes (cspad + other detectors: 3 nodes, two EPIX: 2 nodes). As we run all the data into a single ami session and the best mapping allows max one ami node/dss node, you have less ami power if you have less dss nodes.

One of my DAQ devices has a problem (damage,....):

use "serverStat <DAQ device alias>" to check on the health of the node. Most likely it is prudent to power-cycle this node. Does this not help, you should power cycle the detector/camera/device itself as well. 

This assumes that the detector has recently worked. Otherwise, if the detector configures, but does produce damage an all events, check the trigger timing settings here:

Detector timing settings

Should your detector not configure, it is either not turned on, needs power cycling or some cable is not patched correctly. Please let CDS folks know which detectors you would like to use so we can test them beforehand.

My ipimb has an issue:

Troubleshooting ipimbs is described on this page:Troubleshooting for Controls IPIMB and Wave8s

My detector runs without errors, but I can't see any beam:

Trusting that there actually are photons on the detector that could be seen, the second point are the timing settings. Find the values you should use here:

Detector timing settings

Go to the EVR configuration (with the exception of pgp triggered devices as the EPIX), if you are using aliases in your DAQ, it should be straightforward to find the right trigger channel. Look at all available EVR cards. For the run trigger setting for e.g. CsPad/Cs140k detectors, it might be best to contact your POC.

My data does not seem to be moving well:

check the status of the data moving here: Data Mover Monitoring, this page can be seen one the main "pswww" page as well. Take out the nodes with the issue, assuming your problem is limited to a single node. If it's wider spread, it might warrant a call.


Rayonix:

In the DAQ, "shutdown" first,  then go to 'edit' -> 'rayonix' .
Make sure binning is what you are going to use to take data
Select on the menu 'Update background on config' and click 'apply'. Run the DAQ for few seconds.
Go back to the 'expert mode' and select on the menu 'keep current background' and click apply.
If this doesn't work go to the terminal where the Rayonix server is running (ask Silke), stop it (CRTL-X) and restart it after about 10 sec (CRTL-R). Restart the DAQ.

Technical note:

"serverStat" at this moment works in all hutches when using the machine name or IP, but the "DAQ alias" interpretation feature might not quite work. We hope to improve on this soon.


Expert Troubleshooting (limited permissions)