Procedure to align the transport using the HeNe and align the IR beam to the HeNe.

Assumption is that the Laser Startup Procedure procedure has already been completed and thus an auto-aligner GUI is already open and running on the laser room cameras.

A) Beam transport alignment: HeNe

  1. Press Stop Alignment (red button) in the auto-aligner. This will stop any scripts running on other machines.
  2. Set camera exposure time for transport cameras by pressing HeNe cameras (see Figure 1).
  3. Open S20 Laser Transport Multi Profile Monitor (FACETHOME→Matlab GUIs→S20 Laser Tools→S20 Laser Transport Multi Profile Monitor).
  4. Start display and check profiles look fine (see reference image below), cameras are running, and laser is visible.
  5. Verify ROIs are set somewhat tightly (see above image), if the full camera ROI is visible see Procedures to Fix Common Issues.
  6. If the laser is not visible on B4, B5, or B6
    1. Tick boxes in the auto-aligner for HeNe Near and Far, B0 and B1, and B2 and B3.
    2. Click Clear References then Grab References. This might take a few seconds. A message will be printed in the log saying grabbing is completed.
    3. Click Start Alignment. The software will loop through the list of cameras (top to bottom), calculate offsets and then move motors.
    4. Once the laser is visible on B4, B5, and B6, click Stop Alignment (red button).
  7. Tick boxes for the transport cameras: HeNe Near and Far, B0 and B1, B2 and B3, B4, B5, and, B6.
    1. At this point, all the boxes in the auto-aligner should be checked.
  8. Click Clear References, then Grab References. This might take a few seconds. A message will be printed in the log saying grabbing is completed.
  9. Click Start Alignment. The software will loop through the list of cameras (top to bottom), calculate offsets and then move motors.
  10. Monitor the log to make sure that the offset tolerances were set appropriately and that the auto aligner does its job.
    1. The auto-aligner will write a warning to the log with a reason when it skips an alignment step.
    2. To fix auto-aligner errors, see Procedures to Fix Common Issues.

B) Beam transport alignment: IR

  1. Drive in the compressor beam block, 0 is in (FACETHOME→Experiment→LI20→MOC-LI20-MC03 Channel 3).
  2. Ensure IPOTR foils are out (IPOTR1, IPOTR2 and IPOTR1P, found at FACETHOME→Profile Monitor→LI20→IP1 Table).
  3. If the RP attenuator is not in, set the energy waveplate to 80deg (FACETHOME→Laser→LI20→Infrastructure→XPS moc-la20-ls24 Channel 1).
    1. If the RP attenuator is in, the waveplate should be at maximum energy.
  4. Ensure the EPS shutter is in (FACETHOME→Laser→LI20 Low and High buttons, green means the EPS shutter is in).
  5. Turn LSS-2 key to ON to allow IR beam to go through transport.
  6. On the auto align GUI, press the button IR mode. This will stop the current alignment and opens a new window (see Figure 2).
  7. Use the buttons in the first line to switch from HeNe to IR:
    1. Block HeNe – drives HeNe block to 10 mm (blocked) from 8 mm (open).
    2. Close Iris – sets iris to -5 mm from -40 mm (full beam size).
    3. Set HWP – rotates wave plate 2 to 65 from 50 (s-pol).
    4. Set B0 B1 exposure – sets exposure on B0 and B1 to 1 s (from 1 ms).
    5. Unblock EPS Shutter – opens EPS shutter.
  8. Check that the IR beam profiles look fine on B0 and B1 (see reference image below).
  9. Click Clear IR references, then Grab IR References.
  10. Click Start IR alignment.
  11. Monitor the log to make sure that the offset tolerances were set appropriately and that the auto aligner does its job.
    1. The auto-aligner will write a warning to the log with a reason when it skips an alignment step.
    2. To fix auto-aligner errors, see Procedures to Fix Common Issues.
  12. Once converged, press Abort IR alignment.
  13. Then follow the steps in line 4 to return settings to HeNe alignment mode:
    1. Block EPS Shutter – closes EPS shutter.
    2. Open iris – opens iris from -5 mm to -40 mm (full beam size).
    3. Set HWP – rotates wave plate 2 from 65 to 50 (s-pol).
    4. Set B0 B1 exposure – sets exposure on B0 and B1 to 1 ms (from 1 s).
    5. Unblock HeNe – drives HeNe block from 10 mm (blocked) to 8 mm (open).
    6. Close the IR alignment window.
  14. On main window of auto align GUI, click Clear References, then Grab References.
  15. Then click Start alignment to resume HeNe and laser room auto alignment.
  16. Monitor the log to make sure that the offset tolerances were set appropriately and that the auto aligner does its job.
    1. The auto-aligner will write a warning to the log with a reason when it skips an alignment step.
    2. To fix auto-aligner errors, see Procedures to Fix Common Issues.