Overview
Multi-gain range detectors are more complex than previous LCLS detectors and calibration constants are not calculated by calibman.
Epix10ka
DAQ needs to collect charge-injection xtc file for offsets.
Both dark/offset calibrations do not have to be done for all gain ranges. The calibration system will insert dummy constants for the gain ranges that are not included.
Offsets must be available before pedestals, because pedestal calculation depends on offsets.
epix10ka_offset_calibration
-h
epix10ka_pedestals_calibration
-h
epix10ka_deploy_constants -h
The deploy command must be done after either the offsets or pedestals have changed, or when creating a new experiment.
Jungfrau
jungfrau_ndarr_dark_proc -h
jungfrau_ndarr_dark_proc -d exp=cxi11216:run=
9
,
11
,
12
:smd -s CxiEndstation.
0
:Jungfrau.
0
-n
2000
-u
jungfrau_gain_constants [-h] <=== prints help
jungfrau_gain_constants -d exp=xcsls3716:run=631 -D [-s Jungfrau.0] <=== generate and deploy constants