Camera Hardware
FACET-II has 6 ORCA-Flash4.0 V3 Digital CMOS cameras (C13440-20CU):
The camera connects to the server via a USB-to-fiber link:
Server Hardware
We have two Dell R750 servers in B244 rack. They each serve two ORCA cameras.
Server OS
The server runs "CentOS7 Lite"
Code Block |
---|
[flaci@cpu-li20-pm05]$ uname -r 3.10.0-1160.76.1.el7.x86_64 |
Camera Software
We run 'orcaUsbApp' in 'ProfileMonitorApp'.
Current Issues
- Camera software is not properly integrated into ProfileMonitorApp framework. This issue might explain some of the issues listed below.
- Save/restore does not work. It appears that the IOC does not know the path to the data. This might be fixed?
- Ask Jeremy
- Some PVs readback value does not match the set value, and this can cause issues with Matlab HLAs.
- CMOS:LI20:3505:BinX(Y)_RBV
- CMOS:LI20:3505:DataType
- This might be fixed? Ask Jeremy
- ROI doesn't work with profmon_gui or DAQ
- Need intergration into watchdog and emit IOC for new channel 3516
Less Urgent Issues and Feature requests
- The camera serial number is hardcoded in the IOC st.cmd. This means that if you replace the camera, you have to edit the IOC at the line "epicsEnvSet("SER","305053")
Code Block |
---|
sgess@rhel6-64p ProfileMonitorAD-git $ cat iocBoot/sioc-li20-pm05/cam_config.cmd # Camera details # ------------------------------------------------------ epicsEnvSet("PREFIX", "CMOS:LI20") epicsEnvSet("ID" , "3505") epicsEnvSet("AREA" , "LI20") epicsEnvSet("EVENT" , "223") epicsEnvSet("TYPE" , "Int16") epicsEnvSet("FTVL" , "SHORT") epicsEnvSet("XSIZE" , "2040") epicsEnvSet("YSIZE" , "2040") epicsEnvSet("NELEMENTS", "4198400") epicsEnvSet("CAM" , "ORCA") epicsEnvSet("QSIZE" , "10") epicsEnvSet("BLOCKING", "1") epicsEnvSet("NCHANS" , "2040") epicsEnvSet("X_OFF" , "-3247") epicsEnvSet("Y_OFF" , "-2435") epicsEnvSet("RESOLUTION", "9.9") epicsEnvSet("EGU" , "um") epicsEnvSet("SER", "305053") # End of file |
- If the camera fails or becomes disconnected from the IOC, this is not reflected by the "ASYN.CNCT" PV, but it works for the GigE cameras.
- Event2 integration.
Bin X/Y RBVs are purple because they have been manually overwritten.