  1. Set the pulsed drive level to 15%
  2. Run a cavity characterization
    1. Raise an error if the characterization crashed
    2. Raise an error if the loaded Q is not within 2.5e7 and 5.1e7
    3. Raise an error if the cavity scale factor is not within 10 and 125
  3. Push the loaded Q and cavity scale factor
  4. Reset the data decimation  to 255 for both CW and pulsed
  5. Reset the piezo feedback set point to 0
  6. Run the cavity probe Q calculation

Cavity Ramp-Up to ADES

  1. Set the piezo to feedback
  2. Set the RF mode to SELA
  3. If the cavity is off, set the amplitude to 5MV and turn it on
  4. Walk the amplitude in steps of 0.5MV up to 10MV and steps of 0.1MV above 10MV
    1. Raise an error if the quench interlock is latched
  5. Set the RF mode to SELAP

...