Full source code can be found here. It can be run natively on unix systems using the Local PyDM/Simulacrum Setup instructions

...

Full Procedure

SSA Calibration

  1. Set the drive max to the previously saved drive max found during initial cavity commissioning (or 1 for the the harmonic linearizers)
  2. Recursively run an SSA calibration with decreasing drive max until either the calibration succeeds or it hits one of two fault conditions:
    1. The max forward power is less than 3kW
    2. The desired drive max is less than 0.5

...