  1. Run an SSA calibration
  2. Tune a cavity to resonance (a detune of 0Hz)
  3. Characterize the cavity and calculate the probe Q
  4. Ramp the RF to the desired amplitude (ADES)

SSA Calibration

Using a previously saved drive max found during initial cryomodule commissioning as a starting point (or 1 for the harmonic linearizers), it recursively runs an SSA calibration with decreasing drive max until either the calibration succeeds or the max forward power reads as too low (currently defined as less than 3kW).