Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Procedure to align the main laser through the compressor and the IP area.

Assumption is that the Laser Startup Procedure and Transport Auto-alignment Procedure procedures have already been completed and thus the compressor beam block is inserted.

ND filters should already be inserted in the laser if the RP attenuator is out.

A) Prepare the IP for IR

Steps to make sure no cameras are shot out and the probeline is blocked. The laser should start aligned through the transport but blocked by the beam block in the compressor.

  1. Put in ND filter flippers for IPOTR1 , IPOTR1P and IPOTR2 (FACETHOME→Experiment→S20Beckhoff Digital I/O).
    1. Use the toggle switches on the S20 Laser Control Panel to avoid digging through the Beckhoff menus.
    2. IPOTR1: Top row, fourth module from the left (MAKE SURE YOU AREN’T USING IPOTR1PD).
    3. IPOTR2 : Top row, third module from the left .IPOTR1P: Top row, fourth module from the left(C).
  2. Confirm on the cameras that the filter flippers are in by watching features on the images vanish.

  3. Ensure the OTR foils are out. Double check it (FAECTHOME→Profile FACETHOME→Profile Monitor→S20).

  4. Make sure the target mount is out of the way (FACETHOME→Experiment→LI20FACETHOME→Experiment→LI20→MOC-LI20-MC05).
    1. Channel 2 is horizontal is removed at -150.
    2. Channel 1 is vertical and is removed at 80.
  5. Set probe attenuator to minimum (FACETHOME→Experiment→LI20→MOC-LI20-MC01 Channel 2) and block probe beam .

B) Compressor alignment

Align to the compressor near and far field references. (Do we normally do this alignment with RP attenuator in or out?)

  1. (FACETHOME→Experiment→S20Beckhoff Digital I/O).
    1. Attenuator minimum energy is MC01 Channel 2 at 37deg.
    2. Block probe beam with top row, first module from the left (A).
  2. Verify the laser is blocked in the compressor.
    1. Look at the motor position on compressor beam block, 0 is in (FACETHOME→Experiment→LI20→MOC-LI20-MC03 Channel 3).
    2. Check that no beam is visible on CompNear.

B) Compressor alignment

  1. If the RP attenuator
  2. Check irises are all open in the laser room.
  3. Insert ND in the laser room.
    1. Flip the full power laser ND filter into the common beam path of DMNear and DMFar cameras.
    2. Install the full power ND filter before the lens in the InjNear and InjFar camera line.
    3. Install the full power ND filter in the wavefront sensor and MO beam path.
    4. Flip the ND filters in front of both the wavefront sensor and MO camera to the in position.
  4. Insert ND in the tunnel.
    1. CompNear and CompFar
    2. PBNear and PBFar
  5. Open motorized iris.
  6. Verify both the optics mover in the compressor and the optics mover in the PB are removed from the laser path.
  7. Block laser after MPA.
  8. Remove RP attenuator.
  9. Unblock laser after MPA.
  10. Move compressor beamblock out of the beam.
  11. Align to CompNear and CompFar.

C) PB alignment


Once the beam is aligned through the compressor, align through the PB and the bypass line.